{ "updated_at": "2026-11-11", "scope": "对应用户指定的前九类实体设备。状态仅描述已有证据;补齐路径是待执行工作,不计入已完成资产。", "categories": [ { "category_id": 1, "title": "自动细胞计数仪", "reference_model": "Countess 3 FL", "asset_key": "openflexure", "mapping": "OpenFlexure Cell Counter 打印件装配;开放方案替代,非 Countess 3 FL 同型号。", "robot": "已加载 Franka 尝试计数室取放,但未通过接触验证。", "blocker": "原装夹指接近时碰到样品架或照明支座;抬高后又无法形成稳定双指夹持。计数室厚约 4 mm、露出样品架约 2 mm,上方有照明组件。仍缺光电组件及计数功能。", "feasibility": "需尺寸可核验的细指或专用夹具", "next_step": "已检查不同开口、抓取点和高度。最终录像在接近阶段碰到 IlluminationDovetail,另一次碰到 SlideHolder;抬高后未形成双指夹持。需要有尺寸依据的细指/专用夹具及安装 CAD,再验证侧向退出与装回。不能删除碰撞体或忽略接触来标为成功。", "references": [ { "label": "OpenFlexure Cell Counter 固定来源", "url": "https://github.com/engpol/OpenFlexure_Cell_Counter/tree/e751bfee1a2e1265d8244fb2e27c2ccf448ba741" } ] }, { "category_id": 2, "title": "液滴微流控单细胞处理设备", "reference_model": "达普 Galaxy", "asset_key": null, "auxiliary_asset_key": "poseidon", "mapping": "整机未建成;Poseidon 显微支架仅是辅助件,不填补整机缺口。", "robot": "尚不能进行整机机械臂操作。", "blocker": "已有芯片图纸和部分支架/泵 CAD,但缺完整装配、芯片/管路接口、装载机构及部分文件的复用许可确认。", "feasibility": "已有局部材料,需补整套装配", "next_step": "可继续沿 Drop-seq、HyDrop 或 Poseidon 开放方案核实泵、芯片夹具及装载接口。需要可复用的对应部件和明确的整体装配关系,才能构建可操作装置;现有辅助支架不能当作整台设备。", "references": [ { "label": "Poseidon 原始硬件", "url": "https://github.com/pachterlab/poseidon/tree/5a139fed350bbf5d775ffa9650f465e557b6ccb0" }, { "label": "Drop-seq 作者资料", "url": "https://mccarrolllab.org/dropseq/" } ] }, { "category_id": 3, "title": "微孔单细胞处理设备", "reference_model": "Singleron Matrix NEO", "asset_key": null, "mapping": "整机未建成;Seq-Well 是候选开放方案,尚未作为完整设备纳入。", "robot": "尚不能进行整机机械臂操作。", "blocker": "Matrix NEO 公开尺寸存在冲突且缺完整分件;Seq-Well 的四份 DWG 不等于自动化整机,其 CC BY-NC-ND 许可限制改编分享。", "feasibility": "需明确装配、机构与许可", "next_step": "先确认 Seq-Well 等替代方案的适用许可、完整装配和芯片装载/夹持关系,再评估导入及机械臂任务。当前不能直接凭四份参考图纸标为整机完成。", "references": [ { "label": "Seq-Well 作者 CAD 与协议", "url": "https://shaleklab.com/protocol/seq-well-portable-low-cost-rna-sequencing-of-single-cells-at-high-throughput_old-2/" } ] }, { "category_id": 4, "title": "全自动液体处理工作站", "reference_model": "Hamilton STAR", "asset_key": "flex", "mapping": "Opentrons Flex 单通道 5–1000 µL;同类替代,通道数、工位和 API 与 STAR 不同。", "robot": "已验证 Franka 将 MagSep 空架体放到 Flex D2 工位、松手撤离,再抓取并取回;同一连续流程 32/32 通过。", "blocker": "只验证几何工位上的架体放置,不是官方耗材兼容认证。未验证移液、取吸头和液体过程;Flex 内部执行器参数仍待校准。", "feasibility": "工位放置可用,移液仍需耗材及驱动资料", "next_step": "当前 handling 入口可复现架体装载/取回;原始架体及 Flex 碰撞几何保持不变。后续需匹配板/管/吸头的 CAD、工位注册、夹持固定及驱动参数,才能继续验证移液。", "references": [ { "label": "ChemLab 固定源版本", "url": "https://github.com/mbzuai-nlp/ChemLab-Assets/tree/c711fe1a7a9d26a7efd9d63d2bddc0de62e6b65e" }, { "label": "Opentrons 官方文档", "url": "https://docs.opentrons.com/" } ] }, { "category_id": 5, "title": "PCR热循环仪", "reference_model": "Bio-Rad T100", "asset_key": "archicycler", "mapping": "ROCGENE ArchiCycler 96:当前实物对应;历史 CFX Opus 96 单独保留。", "robot": "Franka + 窄指尖以实际接触完成开盖、从支架取单管入孔、松爪稳定及关盖;该连续 Teacher 装载流程通过独立审计。", "blocker": "额外的连续取回/回架扩展未通过(失抓或掌部碰屏),保留失败录像。照片尺寸、质量、摩擦与窄指尖均为未实测工程假设;入孔使用仿真管姿态反馈。未验证真机、视觉策略、热循环、真实锁扣或自由底座滑移。", "feasibility": "开盖—取管入孔—关盖通过 · Teacher", "next_step": "实测前沿净空、开盖力、孔管公差及窄指尖;真机与视觉策略另行验证。", "references": [ { "label": "厂家系列资料", "url": "https://www.rocgene.cn/General-purpose-PCR-instrument/313" } ], "validation_label": "开盖—取管入孔—关盖通过 · Teacher" }, { "category_id": 6, "title": "磁珠纯化模块", "reference_model": "DynaMag-96", "asset_key": "magsep", "mapping": "MagSep Rack 12 管架体;管式部分替代,非 96 孔板式结构。", "robot": "已验证 Franka 抓取 MagSep 空架体、搬到 Flex 工位并释放,再取回;与第 4 类共用一条 32/32 通过的连续流程。", "blocker": "验证的是空架体搬运,尚非试管插装或磁珠纯化。缺尺寸适配已核验的试管和磁铁模型;架体毫米尺度仍为工程推断。", "feasibility": "架体搬运可用,插管需配套模型", "next_step": "可复用现有架体抓取和释放流程。试管插装仍需许可及尺寸明确的 1.5–2.0 mL 试管 CAD,核对孔口配合后验证;磁场、磁珠和液体过程另行实现。", "references": [ { "label": "MagSep 原始架体与规格", "url": "https://github.com/cmmonaco/MagSep-Rack/tree/66bd2302d2f6daa9911eb5fbf7bfe080a079a93e" } ] }, { "category_id": 7, "title": "温控/混匀反应模块", "reference_model": "INHECO CPAC", "asset_key": "ika", "mapping": "IKA KS 4000 i control;加热/混匀类别的部分替代,不覆盖 CPAC 冷却或紧凑占位。", "robot": "已验证完整开盖 → 装载烧杯 → 释放撤离 → 关盖,36/36 通过。", "blocker": "只通过固定布局机械操作。气弹簧力学未标定,平台轨道为虚拟 XY,加热、液体与混匀过程均未实现。", "feasibility": "现有机械流程可用,扩展需资料", "next_step": "现有流程可按脚本复现。继续扩展需实测气弹簧/摩擦参数及真实偏心机构资料;温控、摇动受力和液体过程分别验证。两支气弹簧数量和布置有厂家依据,不能据此认定全部参数已正确标定。", "references": [ { "label": "IKA 固定源模型", "url": "https://huggingface.co/datasets/MBZUAI/ChemLab-Assets/tree/a6c006f921c9bee62f5585dd7f769163a7f51573/assets/expansion_batch03/t4_measure/ika_ks4000" } ] }, { "category_id": 8, "title": "核酸片段分析仪", "reference_model": "Agilent 4200 TapeStation", "asset_key": null, "mapping": "整机未纳入;SCIEX PA 800 Plus 仅为已筛查候选。", "robot": "尚不能进行已验证的整机机械臂操作。", "blocker": "TapeStation 缺适用整机 CAD;PA 800 源模型将真机四连杆外罩简化为单铰链、托盘固定,不满足当前机构正确性要求。", "feasibility": "需正确机构资料或新来源", "next_step": "继续寻找原有四连杆、样品托盘及进样机构的可复用 CAD/装配依据,或选择资料更完整的同类设备。现有手册与单铰链模型不足以确定关节轴点、连杆尺寸和正确装载轨迹。", "references": [ { "label": "TapeStation 官方手册", "url": "https://www.agilent.com/cs/library/usermanuals/public/4200-TapeStation_SystemManual.pdf" }, { "label": "PA 800 候选模型与限制", "url": "https://huggingface.co/datasets/MBZUAI/ChemLab-Assets/tree/a6c006f921c9bee62f5585dd7f769163a7f51573/assets/expansion_batch04/t3_analytical/sciex_pa800" } ] }, { "category_id": 9, "title": "核酸浓度检测仪", "reference_model": "Qubit Flex", "asset_key": "diynafluor", "mapping": "DIYNAFLUOR 打印件装配;单管开放方案替代,非 Qubit Flex 8 管整机。", "robot": "已验证取帽 → 停放释放 → 重新抓取 → 盖回,32/32 通过。", "blocker": "只通过遮光帽操作。未装载样品管;缺光电组件,尚不能完成荧光检测或浓度读数。", "feasibility": "可继续核对并补样品管装载", "next_step": "原始遮光帽和支座已复用成功。下一步按手册的 0.5 mL 薄壁 PCR/Qubit 管要求,取得匹配耗材模型、核对孔口尺寸与姿态,再验证插管和盖帽。测量需另接光学、电子/算法或有依据的过程模型。", "references": [ { "label": "DIYNAFLUOR 原始 STL、STEP 与手册", "url": "https://github.com/traulab/DIYNAFLUOR/tree/2b4b65022160533f55032c1206cdc6eb5fb7e0fc" } ] } ] }