Datasets:
Tasks:
Tabular Classification
Formats:
csv
Languages:
English
Size:
< 1K
Tags:
dietary-supplements
biomarkers
meta-analysis
randomized-controlled-trials
effect-size
laboratory-medicine
License:
Mirror of supplement-lab-effects 1.0.0 (Zenodo doi:10.5281/zenodo.24543303, GitHub caraulani/supplement-lab-effects)
66aa7dc verified Download data/biotin_interference_thresholds.csv from lifefrom/supplement-lab-effects: direct link, hf CLI and curl.
- Browser
- Download file 21.8 kB
-
https://huggingface.co/datasets/lifefrom/supplement-lab-effects/resolve/main/data/biotin_interference_thresholds.csv
- Command line
-
hf download hf://datasets/lifefrom/supplement-lab-effects/data/biotin_interference_thresholds.csv
-
curl -L -o biotin_interference_thresholds.csv https://huggingface.co/datasets/lifefrom/supplement-lab-effects/resolve/main/data/biotin_interference_thresholds.csv
21.8 kB
| row_id,analyte,marker_slug,assay,platform,manufacturer,assay_format,threshold_value,threshold_unit,threshold_basis,bias_at_threshold_pct,direction,tested_concentrations,sample_type,study_design,pmcid,doi,source_label,location,notes,source_id,full_text_read,access_date,evidence_quote,evidence_quote_status | |
| T24,alpha-fetoprotein,afp,Abbott AFP reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T06,alpha-fetoprotein,afp,Elecsys AFP (older generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,400,ng/mL,lowest tested concentration with |bias| >10%,-16,negative,"25, 50, 100, 200, 400, 600, 800, 1000, 1250, 1500 ng/mL","pooled serum (low, medium, high analyte pools)",in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 2; Table 3,"Table 2 has four columns per analyte (biotin-added old reagent, biotin-added new reagent, biotin-depleted old, biotin-depleted new), read from the header 'Biotin-added (OR/NR) | Biotin-depleted (OR/NR)'; this column order is consistent with the authors' text and Table 3. Authors' Table 3 tolerance for old AFP reagent: <400 ng/mL (package insert <180); the discussion text gives <200 ng/mL instead.",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T07,alpha-fetoprotein,afp,Elecsys AFP (newer generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,,ng/mL,,,,"25, 50, 100, 200, 400, 600, 800, 1000, 1250, 1500 ng/mL (Table 2); 500, 1000, 1500, 2000, 2500, 3000 ng/mL (Table 4)",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 2,no >10% bias up to 3000 ng/mL (Table 4). Package insert ≤1200 ng/mL.,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T35,anti-Müllerian hormone,amh,Ansh picoAMH ELISA,Dynex DS2,Ansh Labs (assay); Dynex (platform),ELISA,,ng/mL,,,,"10,000 ng/mL",serum,in vitro spiking study (method evaluation),PMC6804432,10.1016/j.plabm.2019.e00140,"Garnett 2019, Pract Lab Med",Methods; Table 1,"no >10% bias up to 10,000 ng/mL (single level tested; −2.0%).",S21,yes,2026-11-25,"Biotin interference was also verified by spiking with D-biotin (10,000 ng/mL) (Sigma-Aldrich). || D-biotin 10000 ng/mL | 2.09 (14.93) | −2.0%",included | |
| T22,CA 125,ca-125,Abbott CA125 reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T14,CA 125,ca-125,Elecsys CA 125 (older generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,500,ng/mL,lowest tested concentration with |bias| >10%,-23,negative,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,"500 ng/mL was the lowest level tested, so the true onset may be much lower; package insert <70 ng/mL. The CEA and CA125 rows of Table 4 print identical means.",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T23,CA 15-3,ca-15-3,Abbott CA153 reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T15,CA 15-3,ca-15-3,Elecsys CA 15-3 (older generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,500,ng/mL,lowest tested concentration with |bias| >10%,-49,negative,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,"500 ng/mL was the lowest level tested, so the true onset may be much lower; package insert <100 ng/mL.",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T25,CA 19-9,ca-19-9,Abbott CA199 reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T16,CA 19-9,ca-19-9,Elecsys CA 19-9 (older generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,1000,ng/mL,lowest tested concentration with |bias| >10%,-44,negative,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,Package insert <100 ng/mL; this pool stayed within 10% at 500 ng/mL.,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T36,CA 72-4,,"ARCHITECT CA 72-4 (acridinium label, no biotin-streptavidin)",Abbott ARCHITECT i2000SR and i1000SR,Abbott,sandwich (CMIA),,ng/mL,,,,4678 ng/mL,pooled serum spiked to CA 72-4 4.00 and 10.0 U/mL,"in vitro interference study (method evaluation, CLSI EP07-A2)",PMC9879778,10.1016/j.plabm.2023.e00308,"Yanagihara 2023, Pract Lab Med",Table 3,no >10% bias up to 4678 ng/mL (95% CI limits of % difference between −5.3% and 2.9%). The abstract prints '4875 mg/mL'; the table value is used. No LifeFrom marker slug for CA 72-4.,S62,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T21,carcinoembryonic antigen (CEA),cea,Abbott CEA reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T13,carcinoembryonic antigen (CEA),cea,Elecsys CEA (older generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,500,ng/mL,lowest tested concentration with |bias| >10%,-23,negative,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,"500 ng/mL was the lowest level tested, so the true onset may be much lower; package insert <120 ng/mL.",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T26,free PSA,psa-free,Abbott FPSA reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T17,free PSA,psa-free,Elecsys free PSA (newer generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,2000,ng/mL,lowest tested concentration with |bias| >10%,-13,negative,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T31,free T3,free-t3,Dimension Vista FT3 (LOCI),Siemens Dimension Vista 1500,Siemens Healthineers,competitive (LOCI),50,ng/mL,lowest tested concentration with |bias| >10%,,positive,"50, 100, 250, 500, 750, 1000, 1200 ng/mL",pooled serum (leftover specimens),in vitro spiking study,PMC6818550,10.1002/jcla.22667,"Dasgupta 2019, J Clin Lab Anal",Results; Table 1,"50 ng/mL was the lowest level tested. Authors' stated threshold for >20% (clinically significant) difference: 100 ng/mL. FT4 and TSH on the same analyzer: the paper states only the >20% level (500 ng/mL), not the >10% level, so they are not tabulated here.",S14,yes,2026-11-25,,withheld: source licence is no open licence stated; free to read at PubMed Central (publisher copyright); see location | |
| T29,free T4,free-t4,Abbott FT4 reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T01,free T4,free-t4,Elecsys FT4 II (2nd generation),Roche cobas e602,Roche Diagnostics,competitive,50,ng/mL,lowest tested concentration with |bias| >10%,,positive,"12.5, 25, 50, 100, 200, 400 ng/mL","pooled human serum (low, normal and high FT4 pools)",in vitro spiking study,PMC6933059,10.3343/alm.2020.40.3.274,"Choi 2020, Ann Lab Med",Text (results paragraph); Table 1,"Lowest level across the three pools; normal-FT4 pool first exceeded 10% at 100 ng/mL. Table 1 (table text present only in the article XML, not in the rendered .txt) gives 12.2% (low pool) and 14.6% (high pool) at 50 ng/mL and 9.7% (normal pool). Data sheet threshold quoted by the authors: ≤25 ng/mL.",S10,yes,2026-11-25,,withheld: source licence is CC BY-NC 4.0; see location | |
| T02,free T4,free-t4,Elecsys FT4 III (3rd generation),Roche cobas e602,Roche Diagnostics,competitive,200,ng/mL,lowest tested concentration with |bias| >10%,39.2,positive,"12.5, 25, 50, 100, 200, 400 ng/mL","pooled human serum (low, normal and high FT4 pools)",in vitro spiking study,PMC6933059,10.3343/alm.2020.40.3.274,"Choi 2020, Ann Lab Med",Text (results paragraph); Table 1,200 ng/mL is the high-FT4 pool; low and normal pools first exceeded 10% at 400 ng/mL. Table 1 (article XML) shows all pools within 10% up to 100 ng/mL. Revised manufacturer threshold quoted by the authors: ≤100 ng/mL.,S10,yes,2026-11-25,,withheld: source licence is CC BY-NC 4.0; see location | |
| T08,free T4,free-t4,Elecsys FT4 (older generation reagent),Roche Cobas e602,Roche Diagnostics,competitive,400,ng/mL,lowest tested concentration with |bias| >10%,76,positive,"25, 50, 100, 200, 400, 600, 800, 1000, 1250, 1500 ng/mL","pooled serum (low, medium, high analyte pools)",in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 2; Table 3,"Old FT4 reagent read −9% at 25, −10% at 50 and +10% at 200 ng/mL (none above 10% in absolute value), then +76% at 400. Authors' Table 3 tolerance: <200 ng/mL; the discussion text gives <50 ng/mL. The paper is internally inconsistent here. Table 2 has four columns per analyte (biotin-added old reagent, biotin-added new reagent, biotin-depleted old, biotin-depleted new), read from the header 'Biotin-added (OR/NR) | Biotin-depleted (OR/NR)'; this column order is consistent with the authors' text and Table 3.",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T09,free T4,free-t4,Elecsys FT4 (newer generation reagent),Roche Cobas e602,Roche Diagnostics,competitive,,ng/mL,,,,"25, 50, 100, 200, 400, 600, 800, 1000, 1250, 1500 ng/mL (Table 2); 500, 1000, 1500, 2000, 2500, 3000 ng/mL (Table 4)",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 2,no >10% bias up to 3000 ng/mL (Table 4). Package insert ≤1200 ng/mL.,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T19,hepatitis B e antigen (HBeAg),,Elecsys HBeAg (older generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich (qualitative assay; change in signal value),500,ng/mL,lowest tested concentration with |bias| >10%,52,positive,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,"Rise rather than fall, as printed. 500 ng/mL was the lowest level tested. No LifeFrom marker slug for HBeAg.",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T20,hepatitis B surface antigen (HBsAg),hbsag,Elecsys HBsAg (newer generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,1000,ng/mL,lowest tested concentration with |bias| >10%,-12,negative,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,Authors note HBsAg tolerance was lower than the manufacturer claim.,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T33,high-sensitivity cardiac troponin I,troponin-i-hs,Atellica IM High-Sensitivity Troponin I (TnIH),Siemens Atellica IM,Siemens Healthineers,sandwich (biotinylated antibodies pre-bound to streptavidin solid phase),,µg/L,,,,"100, 1500 µg/L",serum at troponin I 50 and 100 ng/L,in vitro spiking study (method evaluation),PMC9195607,10.11613/bm.2022.020709,"González de la Presa 2022, Biochem Med (Zagreb)",Results; Table 4,no >10% bias up to 1500 µg/L; largest bias −1.67%.,S25,yes,2026-11-25,"Serum samples were spiked with biotin stock to achieve biotin concentrations of 100 µg/L and 1500 µg/L. || No interference was observed for biotin ≤ 1,500 µg/L. || Spike (1500) | 100 | 92.455 | -1.67",included | |
| T03,high-sensitivity cardiac troponin T,troponin-t-hs,Elecsys hs-cTnT,Roche Cobas 8000 module e602,Roche Diagnostics,sandwich,50,µg/L,lowest tested concentration with |bias| >10%,-23.5,negative,"50, 100, 500, 1000 µg/L","pooled lithium-heparin plasma at four troponin T levels (18, 59, 201, 6423 ng/L)",in vitro spiking study,PMC6214697,10.11613/bm.2018.030901,"Schrapp 2018, Biochem Med (Zagreb)",Results; Table 1,"50 µg/L was the lowest level tested, so the true onset may be lower. Per-pool negative biases at 50 µg/L ranged 21.6-25.4% (Table 1). Bias is a fall in measured troponin T.",S50,yes,2026-11-25,"We observed a mean negative bias of initial troponin concentration of 23.5%, 55.8%, 96.7%, 98.2% at 50, 100, 500, 1000 µg/L biotin concentrations, respectively || Mean of biases | 23.5% | 55.8% | 96.7% | 98.2%",included | |
| T34,holotranscobalamin (Active B12),holotranscobalamin,Elecsys Active B12,Roche cobas e immunoassay analyzers (multicentre),Roche Diagnostics,Elecsys ECLIA (biotin-streptavidin),,ng/mL,,,,dilution series from a serum sample with 100 ng/mL biotin (individual levels not listed); highest level reported without interference 40 ng/mL,"serum at active B12 34.1, 70.4 and 120 pmol/L",in vitro interference study (method evaluation),PMC6826884,10.1177/0004563219863818,"Heil 2019, Ann Clin Biochem",Methods; Results,"no >10% bias up to 40 ng/mL (acceptance: ≤10% for 8-100 pmol/L, ≤15% above 100, ≤0.8 pmol/L absolute for 3-8 pmol/L). The paper does not say whether levels between 40 and 100 ng/mL exceeded the limit.",S28,yes,2026-11-25,,withheld: source licence is CC BY-NC 4.0; see location | |
| T32,NT-proBNP,nt-probnp,Architect NT-proBNP,Abbott Architect i2000,Abbott,sandwich (CMIA),,ng/mL,,,,500 to 4250 ng/mL,serum at two medical decision levels (80 and 1800 pg/mL),in vitro spiking study (method evaluation),PMC9140882,10.3390/diagnostics12051172,"Lau 2022, Diagnostics (Basel)",Methods; Results; Table 2,no >10% bias up to 4250 ng/mL; largest change ≤3.07%. Manufacturer claim quoted by the authors: 4250 ng/mL.,S32,yes,2026-11-25,spiked into serum samples at two Medical Decision Levels (MDLs: 80 pg/mL and 1800 pg/mL) at a final concentration of 500 to 4250 ng/mL || The percentage bias between a control sample and a sample spiked with levels as high as 4250 ng/mL biotin met the acceptance criteria of ≤10%,included | |
| T04,thyroglobulin,thyroglobulin,Elecsys Tg (older generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,50,ng/mL,lowest tested concentration with |bias| >10%,-19,negative,"25, 50, 100, 200, 400, 600, 800, 1000, 1250, 1500 ng/mL","pooled serum (low, medium, high analyte pools)",in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 2; Table 3,"Table 2 has four columns per analyte (biotin-added old reagent, biotin-added new reagent, biotin-depleted old, biotin-depleted new), read from the header 'Biotin-added (OR/NR) | Biotin-depleted (OR/NR)'; this column order is consistent with the authors' text and Table 3. Authors' Table 3 tolerance for old Tg reagent: <50 ng/mL (package insert <30).",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T05,thyroglobulin,thyroglobulin,Elecsys Tg (newer generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,2000,ng/mL,lowest tested concentration with |bias| >10%,-12,negative,"25, 50, 100, 200, 400, 600, 800, 1000, 1250, 1500 ng/mL (Table 2); 500, 1000, 1500, 2000, 2500, 3000 ng/mL (Table 4)",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 2,"Table 2 (separate experiment, levels to 1500 ng/mL): new Tg reagent stayed within 10% (−8% at 1500). Table 4 extends to 3000 ng/mL; first >10% at 2000. Package insert ≤1200 ng/mL. Table 2 has four columns per analyte (biotin-added old reagent, biotin-added new reagent, biotin-depleted old, biotin-depleted new), read from the header 'Biotin-added (OR/NR) | Biotin-depleted (OR/NR)'; this column order is consistent with the authors' text and Table 3.",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T28,thyroglobulin antibody (anti-Tg),thyroglobulin-antibody,Abbott ATG reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T10,thyroglobulin antibody (anti-Tg),thyroglobulin-antibody,Elecsys Anti-Tg (older generation reagent),Roche Cobas e602,Roche Diagnostics,competitive,50,ng/mL,lowest tested concentration with |bias| >10%,13,positive,"25, 50, 100, 200, 400, 600, 800, 1000, 1250, 1500 ng/mL","pooled serum (low, medium, high analyte pools)",in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 2; Table 3,"Table 2 has four columns per analyte (biotin-added old reagent, biotin-added new reagent, biotin-depleted old, biotin-depleted new), read from the header 'Biotin-added (OR/NR) | Biotin-depleted (OR/NR)'; this column order is consistent with the authors' text and Table 3. Authors' Table 3 tolerance: <50 ng/mL (package insert <60).",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T11,thyroglobulin antibody (anti-Tg),thyroglobulin-antibody,Elecsys Anti-Tg (newer generation reagent),Roche Cobas e602,Roche Diagnostics,competitive,1250,ng/mL,lowest tested concentration with |bias| >10%,13,positive,"25, 50, 100, 200, 400, 600, 800, 1000, 1250, 1500 ng/mL (Table 2); 500, 1000, 1500, 2000, 2500, 3000 ng/mL (Table 4)",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 2; Table 3; Table 4,"Table 4 (separate pool, 500-3000 ng/mL) first exceeded 10% at 1500 ng/mL (11%). Package insert ≤1200 ng/mL. Table 2 has four columns per analyte (biotin-added old reagent, biotin-added new reagent, biotin-depleted old, biotin-depleted new), read from the header 'Biotin-added (OR/NR) | Biotin-depleted (OR/NR)'; this column order is consistent with the authors' text and Table 3.",S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T30,thyrotropin (TSH),tsh,Abbott TSH reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T12,thyrotropin (TSH),tsh,Elecsys TSH (newer generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,2000,ng/mL,lowest tested concentration with |bias| >10%,-23,negative,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,Package insert ≤1200 ng/mL; authors' Table 6 tolerance <1500 ng/mL. Lowest level tested was 500 ng/mL.,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T27,total PSA,psa-total,Abbott TPSA reagent (as labelled in Table 5),Abbott Architect i2000SR,Abbott,CMIA (no biotin-streptavidin capture),,ng/mL,,,,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 5,no >10% bias up to 3000 ng/mL,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |
| T18,total PSA,psa-total,Elecsys total PSA (newer generation reagent),Roche Cobas e602,Roche Diagnostics,sandwich,3000,ng/mL,lowest tested concentration with |bias| >10%,-13,negative,"500, 1000, 1500, 2000, 2500, 3000 ng/mL",pooled serum,in vitro spiking study,PMC12099460,10.1016/j.plabm.2025.e00472,"Chiu 2025, Pract Lab Med",Table 4; Table 6,Only the highest level tested exceeded 10%.,S09,yes,2026-11-25,,withheld: source licence is CC BY-NC-ND 4.0; see location | |