name: twente-vortexflow
pretty_name: >-
  Flow Phantom Ultrasound Channel/Optical Data (Physics of Fluids, University of
  Twente)
license: cc-by-4.0
task_categories:
  - image-segmentation
  - other
tags:
  - ultrasound
  - rf
  - open-h-rf
  - flow
  - phantom
  - blood-flow
language:
  - en
size_categories:
  - 1K<n<10K

Twente Ultrasound-Optical Flow Phantom Chamber Data

Optical camera view beside the B-mode reconstruction of a von Karman vortex street

The Photron high-speed camera view (left) and the B-mode reconstruction (right) of data/AcqData_PVoltage80_TVoltage3.4.hdf5. Both come from track_0, frame for frame, so the optical and acoustic views show the same instant of the vortex street.

Dataset Description

Pre-beamformed ultrasound channel-capture data acquired with a curved-array transducer (GEC1-6D, 192 elements, 3.4 MHz center frequency) from a flow phantom, accompanied with simultaneously recorded camera images. The phantom contains a flow chamber through which a water with optical and acoustical scatterers is pumped at controlled flow rates. Six acquisitions are provided, spanning three pump voltage levels (80 V, 120 V, 160 V) and two transmit voltage levels (3.4 V, 7.1 V), each capturing two transmit types: a short imaging pulse and a chirp waveform. Each acquisition contains 750 frames of single plane-wave RF channel data. The intended task is blood-flow imaging and Doppler processing (RFP task group 6.2).

Phantom

The front and the back of the flow chamber are made from medical-grade gelatin to facilitate ultrasound transmission. A cylinder with a diameter of 6 mm is placed inside the flow chamber which generates a von Kármán vortex street. The distance between the walls of the flow chamber is about 3 cm. A schematic of the setup is shown in Figure 1. Figure 1: Ultrasound - optical flow phantom setup

The elevation focus of the transducer is aligned with the optical light sheet, see Figure 2. Figure 2: Alignment of the acoustical beam and the lightsheet

Contrast

Optical scattering was facilitated by hollow glass beads (mean particle size: 9-13 micrometer, Manufacturer: Sigma-Aldrich, PubChem Substance ID: 24867590). The acoustical scatter was enhanced by adding in-house produced microbubbles. The microbubble size distribution is shown in Figure 3. Figure 3: Microbubble size distribution

Acquisition parameters

The acquisition settings for all six datasets are summarized in Table 1.

Table 1. Acquisition parameters per dataset.

Dataset number Dataset name Transformer output (V) Est. pump output (L/s) Transducer driving (V)
1 AcqData_PVoltage80_TVoltage3.4 80 0.056 3.4
2 AcqData_PVoltage80_TVoltage7.1 80 0.056 7.1
3 AcqData_PVoltage120_TVoltage3.4 120 0.107 3.4
4 AcqData_PVoltage120_TVoltage7.1 120 0.107 7.1
5 AcqData_PVoltage160_TVoltage3.4 160 0.138 3.4
6 AcqData_PVoltage160_TVoltage7.1 160 0.138 7.1

Dataset Contributor(s)

  • Rienk Zorgdrager r.c.zorgdrager@utwente.nl (ORCiD: 0009-0001-2537-117X)
  • Guillaume Lajoinie
  • Michel Versluis
  • Physics of Fluids Group, Faculty of Science and Technology, University of Twente

Dataset Creation Date

07/01/2026

License / Terms of Use

Creative Commons Attribution 4.0 International (CC BY 4.0). Retain attribution and identify modifications when reusing the data.

Intended Usage

Suitable for research in:

  • Ultrasound localization microscopy (ULM) / super-resolution flow imaging
  • Validation/verification of ultrasound flow imaging techniques using optical references
  • Fluid dynamics using ultrasound
  • Chirp compression and coded-excitation beamforming
  • Beamforming quality comparison across transmit voltage levels (SNR studies)

Dataset Characterization

  • Data Collection Method: Phantom / table-top (flow phantom, no human subjects)
  • Labeling Method: No manual labels; ground-truth flow rate is implicit in camera images. Note that the measured velocity may differ from the pump output in Table 1 due to changes in geometry and flow profiles in the flow chamber.
  • Acquisition system:
    • Transducer: GEC1-6D curved array, 192 elements, 3.4 MHz center frequency, 95% bandwidth, 35 µm element width, 66 mm elevation focus, 0.0568 m radius
    • Transmit: diverging wave (no transmit delays, so the diverging nature is induced by the curvature of the surface)
    • Sampling rate: ~19.2 MHz
    • Sound speed used: 1509.6 m/s (water-based phantom)
    • Data type: raw RF (n_ch = 1, float32)
    • System: Verasonics Vantage 256

Processing the Dataset

The acquisitions can be processed with the reconstruct.py script as provided in the Open-H-RF GitHub repository, together with the pipeline definitions in this folder and the zea library. The script streams the data from the Hugging Face Hub.

ZEA_FILE and FRAME at the top of the script select the acquisition and frame; each track is reconstructed with its own pipeline (pipeline_short_imaging_pulse.yaml, pipeline_chirp.yaml).

Dataset Format

zea v0.1.6

All files are in the zea format (HDF5 + zea schema, current release zea_version 0.1.6). Each .hdf5 file contains two tracks:

Table 2. Track labels.

Track label Description
short imaging pulse Standard narrow-band pulse transmit
chirp Frequency-swept (chirp) coded excitation transmit

Both tracks use the same probe and geometry. The raw channel data arrays are stored as float32 and are pre-beamformed (not yet envelope-detected or log-compressed).

No pre-processing (demodulation, decimation, filtering) has been applied before packaging.

Dataset Quantification

Current Open-H-RF release: 6 HDF5 files; 9.09 GB (9,088,991,232 bytes) stored; root zea_version 0.1.6. Sizes include all HDF5 contents and use decimal units (MB = 10^6 bytes, GB = 10^9 bytes, TB = 10^12 bytes), not decoded-array memory or original-source download sizes.

Table 3. Acquisition settings.

File Pump V TX V Frames per track Tracks Stored HDF5 size
AcqData_PVoltage80_TVoltage3.4.hdf5 80 V 3.4 V 750 2 1.44 GB
AcqData_PVoltage80_TVoltage7.1.hdf5 80 V 7.1 V 750 2 1.55 GB
AcqData_PVoltage120_TVoltage3.4.hdf5 120 V 3.4 V 750 2 1.50 GB
AcqData_PVoltage120_TVoltage7.1.hdf5 120 V 7.1 V 750 2 1.54 GB
AcqData_PVoltage160_TVoltage3.4.hdf5 160 V 3.4 V 750 2 1.51 GB
AcqData_PVoltage160_TVoltage7.1.hdf5 160 V 7.1 V 750 2 1.55 GB

Total frames: 9,000 (6 files × 750 frames), each covering 2 transmit types.

  • Stored HDF5 size: 9.09 GB (9,088,991,232 bytes). No train/validation/test split is defined; all acquisitions are provided as-is.

Per-sample feature table

Table 4. Per-sample features.

Name Shape (per frame) Dtype Units Description
raw_data (1, 3456, 192, 1) float32 — Pre-beamformed RF channel data (1 plane-wave TX)
image/values (646, 435) uint8 — Pre-computed B-mode image (stored in file, uint8)
scan/sampling_frequency scalar float32 Hz A/D sampling rate (~19.2 MHz)
scan/sound_speed scalar float32 m/s Speed of sound used for reconstruction (~1509.6 m/s)
scan/t0_delays (1, 192) float32 s Per-element transmit delays (plane-wave: all zeros)
scan/tx_apodizations (1, 192) float32 — Transmit apodization (all ones = uniform)
scan/tgc_gain_curve (3456,) float32 dB Time-gain compensation curve
probe/probe_geometry (192, 3) float32 m Element positions (x, y, z) in metres

Subject Metadata

This is a phantom dataset (no human or animal subjects). Flow rates are controlled by pump voltage (80 V, 120 V, 160 V), see Table 1.

Data Validation

reconstruct.py uses one pipeline YAML file per track:

  • pipeline_short_imaging_pulse.yaml — for the short imaging pulse track
  • pipeline_chirp.yaml — for the chirp track

The pipeline applies: Cast(float32) → Demodulate → Beamform(DAS, 100 patches) → EnvelopeDetect → Normalize → LogCompress

Reference B-mode image (AcqData_PVoltage80_TVoltage3.4.hdf5, frame 10):

Reference B-mode reconstruction

Top: short imaging pulse track. Bottom: chirp track. Two horizontal phantom wall reflections are visible, with a speckle-filled flow chamber between them. Near-field reverberation and grating-lobe artifacts at the walls and the cylinder are acquisition-induced.

Reference mapping between camera and ultrasound image (AcqData_PVoltage80_TVoltage3.4.hdf5, frame 10):

Reference images of particles in flow Top: short imaging pulse track. Bottom: synchronized camera recording. The walls of the phantom and the cylinder are visible in both images. In the ultrasound image, speckle is visible in between the walls (mainly bubble induced), whereas in the camera image the contrast is induced by the hollow glass beads. Light reflection artefacts are visible in the camera image near the cylinder and the walls.

Known Issues

  • The ultrasound recordings made with the chirp contain clipped reflections at the interface between walls and the water.
  • The speed of sound differs between the water and the medical gelatin (1449.30 +- 3.37 m/s).
  • A chirp compression algorithm is not provided.
  • The center frequency of the chirp is determined as the mean of the input frequency for the associated cycle in the Verasonics. This may therefore only be considered a very rough estimation.
  • An image registration algorithm is not provided, but the camera pixel size can be estimated using the geometry of the flow chamber.

Ethical Considerations

This is a phantom dataset with no human or animal subjects. No IRB approval or informed consent is required. No personally identifiable information is present.

The phantom and flow phantom components do not carry proprietary IP constraints.