MesserMMP's picture
Rebuild CardioSyntax with custom bilingual UI and verified inference
1024e04 verified
Raw History Blame Contribute Delete
4.05 kB
"""DICOM input checks; no source files or patient metadata are modified."""
from pathlib import Path
MIN_FRAMES = 10
MAX_FRAMES = 1024
MAX_DECODED_BYTES = 512 * 1024 * 1024
class InferenceError(RuntimeError):
"""An actionable validation or inference failure, never a numeric fallback."""
def inspect(path):
import pydicom
tags = ["StudyInstanceUID", "SeriesInstanceUID", "SOPInstanceUID", "NumberOfFrames",
"Rows", "Columns", "SamplesPerPixel", "BitsAllocated", "PixelRepresentation",
"PhotometricInterpretation"]
try:
ds = pydicom.dcmread(str(path), stop_before_pixels=True, specific_tags=tags)
result = {
"path": str(path),
"study_uid": str(ds.get("StudyInstanceUID", "")),
"series_uid": str(ds.get("SeriesInstanceUID", "")),
"sop_uid": str(ds.get("SOPInstanceUID", "")),
"frames": int(ds.get("NumberOfFrames", 1)),
"rows": int(ds.get("Rows", 0)),
"columns": int(ds.get("Columns", 0)),
"samples_per_pixel": int(ds.get("SamplesPerPixel", 1)),
"bits_allocated": int(ds.get("BitsAllocated", 0)),
"pixel_representation": int(ds.get("PixelRepresentation", 0)),
"photometric_interpretation": str(ds.get("PhotometricInterpretation", "")),
}
except Exception as exc:
raise InferenceError(f"Cannot read DICOM header: {Path(path).name}.") from exc
result["eligible"] = (
MIN_FRAMES <= result["frames"] <= MAX_FRAMES
and result["rows"] > 0 and result["columns"] > 0
and result["samples_per_pixel"] == 1
and result["bits_allocated"] in (8, 16)
and result["pixel_representation"] == 0
and result["photometric_interpretation"] in ("MONOCHROME1", "MONOCHROME2")
)
return result
def read_video(path):
"""Return raw unsigned grayscale frames; preserve the reference pixel convention."""
import numpy as np
import pydicom
meta = inspect(path)
if not meta["eligible"]:
raise InferenceError(
f"{Path(path).name}: use an unsigned 8/16-bit grayscale DICOM cine "
f"with {MIN_FRAMES}–{MAX_FRAMES} frames."
)
if meta["frames"] * meta["rows"] * meta["columns"] * (meta["bits_allocated"] // 8) > MAX_DECODED_BYTES:
raise InferenceError(f"{Path(path).name}: the decoded video exceeds the memory limit.")
try:
video = pydicom.dcmread(str(path)).pixel_array
except Exception as exc:
raise InferenceError(f"Cannot decode DICOM pixels: {Path(path).name}.") from exc
if video.ndim != 3 or tuple(video.shape) != (meta["frames"], meta["rows"], meta["columns"]):
raise InferenceError(f"{Path(path).name}: pixel dimensions do not match its DICOM header.")
if video.dtype.kind != "u" or video.dtype.itemsize not in (1, 2):
raise InferenceError(f"{Path(path).name}: unsupported pixel data type.")
# Rescale/windowing and MONOCHROME1 inversion are deliberately not applied:
# the published model reads pixel_array directly using these conventions.
return np.asarray(video)
def validate_study(paths):
"""Reject mixed studies and duplicate acquisitions without reordering paths."""
records = [inspect(path) for path in paths]
if not records:
raise InferenceError("Select at least one angiography projection.")
if any(not record["study_uid"] or not record["sop_uid"] for record in records):
raise InferenceError("Each DICOM must retain StudyInstanceUID and SOPInstanceUID.")
if len({r["study_uid"] for r in records}) != 1:
raise InferenceError("The selected projections belong to different studies.")
if len({r["sop_uid"] for r in records}) != len(records):
raise InferenceError("A projection is included more than once (duplicate SOPInstanceUID).")
if any(not r["eligible"] for r in records):
raise InferenceError("Remove unsupported or fewer-than-10-frame projections before inference.")
return records