  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
15
  </header>
16
  <h2 id="context-of-use-cou">Context of Use (COU)</h2>
 
8
  </head>
9
 
10
  <header>
 
 
 
11
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
12
  </header>
13
  <h2 id="context-of-use-cou">Context of Use (COU)</h2>
static/COU.md CHANGED
@@ -9,9 +9,6 @@
9
  </head>
10
 
11
  <header>
12
- <img src="images/FDAlogo.png" style="float: left;" height="100"/>
13
- <img src="images/FDAgraphic.png" style="float: right;" height="100"/>
14
- <br clear="all" />
15
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
16
  </header>
17
  ```
 
9
  </head>
10
 
11
  <header>
 
 
 
12
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
13
  </header>
14
  ```
static/FAQ.html CHANGED
@@ -7,9 +7,6 @@
7
  </head>
8
 
9
  <header>
10
- <img src="images/FDAlogo.png" style="float: left;" height="100"/>
11
- <img src="images/FDAgraphic.png" style="float: right;" height="100"/>
12
- <br clear="all" />
13
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
14
  </header>
15
  <h2 id="frequently-asked-questions">Frequently asked questions</h2>
 
7
  </head>
8
 
9
  <header>
 
 
 
10
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
11
  </header>
12
  <h2 id="frequently-asked-questions">Frequently asked questions</h2>
static/FAQ.md CHANGED
@@ -8,9 +8,6 @@
8
  </head>
9
 
10
  <header>
11
- <img src="images/FDAlogo.png" style="float: left;" height="100"/>
12
- <img src="images/FDAgraphic.png" style="float: right;" height="100"/>
13
- <br clear="all" />
14
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
15
  </header>
16
  ```
 
8
  </head>
9
 
10
  <header>
 
 
 
11
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
12
  </header>
13
  ```
static/README.html CHANGED
@@ -7,14 +7,10 @@
7
  </head>
8
 
9
  <header>
10
- <img src="images/FDAlogo.png" style="float: left;" height="100"/>
11
- <img src="images/FDAgraphic.png" style="float: right;" height="100"/>
12
- <br clear="all" />
13
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
14
  </header>
15
  <h2 id="overview">Overview</h2>
16
  <p>The Color Hazard and RISk calculator (CHRIS) is a tool under development within the Center for Devices and Radiological Health (CDRH) at the Food and Drug Administration (FDA). The tool enables the user to perform rapid, screening level toxicological risk assessments of color additives and additives and impurities associated with the color additive that might be released from a polymer medical device component. CHRIS applies a model that computes an exposure estimate and compares the result to a tolerable exposure (TE) or threshold of toxicological concern (TTC) value. The tool assists the user in determining (a) whether the toxicological risk of a color additive associated substance should be further assessed and/or (b) the potential impact of the substance on device biocompatibility.</p>
17
- <p><strong>Disclaimer: CHRIS only addresses risk associated with the presence of color additives and additives and impurities associated with the color additive. Therefore, a favorable outcome by CHRIS does not imply acceptable biological risk for the final finished form of a medical device. Further, a successful CHRIS evaluation does not address local biocompatibility endpoints, and therefore additional testing may be needed despite a positive outcome from CHRIS. Because there may be additional restrictions based on device specific guidance and code of federal regulation (CFR) listings, users should also review the relevant documents for their device type and color additive.</strong></p>
18
  <h2 id="important-considerations">Important considerations</h2>
19
  <ol type="1">
20
  <li><p>CHRIS assists in screening the potential risk of a color additive associated substance in a polymer medical device component. Because color additives and concentrates may contain multiple distinct substances, e.g. impurities and carrier resins, CHRIS can be used to evaluate each substance present in the component individually.</p></li>
 
7
  </head>
8
 
9
  <header>
 
 
 
10
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
11
  </header>
12
  <h2 id="overview">Overview</h2>
13
  <p>The Color Hazard and RISk calculator (CHRIS) is a tool under development within the Center for Devices and Radiological Health (CDRH) at the Food and Drug Administration (FDA). The tool enables the user to perform rapid, screening level toxicological risk assessments of color additives and additives and impurities associated with the color additive that might be released from a polymer medical device component. CHRIS applies a model that computes an exposure estimate and compares the result to a tolerable exposure (TE) or threshold of toxicological concern (TTC) value. The tool assists the user in determining (a) whether the toxicological risk of a color additive associated substance should be further assessed and/or (b) the potential impact of the substance on device biocompatibility.</p>
 
14
  <h2 id="important-considerations">Important considerations</h2>
15
  <ol type="1">
16
  <li><p>CHRIS assists in screening the potential risk of a color additive associated substance in a polymer medical device component. Because color additives and concentrates may contain multiple distinct substances, e.g. impurities and carrier resins, CHRIS can be used to evaluate each substance present in the component individually.</p></li>
static/README.md CHANGED
@@ -8,9 +8,6 @@
8
  </head>
9
 
10
  <header>
11
- <img src="images/FDAlogo.png" style="float: left;" height="100"/>
12
- <img src="images/FDAgraphic.png" style="float: right;" height="100"/>
13
- <br clear="all" />
14
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
15
  </header>
16
  ```
@@ -19,8 +16,6 @@
19
 
20
  The Color Hazard and RISk calculator (CHRIS) is a tool under development within the Center for Devices and Radiological Health (CDRH) at the Food and Drug Administration (FDA). The tool enables the user to perform rapid, screening level toxicological risk assessments of color additives and additives and impurities associated with the color additive that might be released from a polymer medical device component. CHRIS applies a model that computes an exposure estimate and compares the result to a tolerable exposure (TE) or threshold of toxicological concern (TTC) value. The tool assists the user in determining (a) whether the toxicological risk of a color additive associated substance should be further assessed and/or (b) the potential impact of the substance on device biocompatibility.
21
 
22
- **Disclaimer: CHRIS only addresses risk associated with the presence of color additives and additives and impurities associated with the color additive. Therefore, a favorable outcome by CHRIS does not imply acceptable biological risk for the final finished form of a medical device. Further, a successful CHRIS evaluation does not address local biocompatibility endpoints, and therefore additional testing may be needed despite a positive outcome from CHRIS. Because there may be additional restrictions based on device specific guidance and code of federal regulation (CFR) listings, users should also review the relevant documents for their device type and color additive.**
23
-
24
  ## Important considerations
25
 
26
  (1) CHRIS assists in screening the potential risk of a color additive associated substance in a polymer medical device component. Because color additives and concentrates may contain multiple distinct substances, e.g. impurities and carrier resins, CHRIS can be used to evaluate each substance present in the component individually.
 
8
  </head>
9
 
10
  <header>
 
 
 
11
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) </h1>
12
  </header>
13
  ```
 
16
 
17
  The Color Hazard and RISk calculator (CHRIS) is a tool under development within the Center for Devices and Radiological Health (CDRH) at the Food and Drug Administration (FDA). The tool enables the user to perform rapid, screening level toxicological risk assessments of color additives and additives and impurities associated with the color additive that might be released from a polymer medical device component. CHRIS applies a model that computes an exposure estimate and compares the result to a tolerable exposure (TE) or threshold of toxicological concern (TTC) value. The tool assists the user in determining (a) whether the toxicological risk of a color additive associated substance should be further assessed and/or (b) the potential impact of the substance on device biocompatibility.
18
 
 
 
19
  ## Important considerations
20
 
21
  (1) CHRIS assists in screening the potential risk of a color additive associated substance in a polymer medical device component. Because color additives and concentrates may contain multiple distinct substances, e.g. impurities and carrier resins, CHRIS can be used to evaluate each substance present in the component individually.
templates/index.html CHANGED
@@ -11,23 +11,13 @@
11
 
12
  </head>
13
 
14
- <img src="{{ url_for('static',filename='images/FDAlogo.png') }}" style="float: left;" height="100"/>
15
- <img src="{{ url_for('static',filename='images/FDAgraphic.png') }}" style="float: right;" height="100"/>
16
- <br clear="all" />
17
-
18
  <header>
19
- <h1 style="text-align:center">CHRIS: FDA's <font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator</h1>
20
  </header>
21
 
22
- <p><strong>Disclaimer: CHRIS only addresses risk associated with the presence of color additives and additives and impurities
23
- associated with the color additive. Therefore, a favorable outcome by CHRIS does not imply acceptable biological risk for the
24
- final finished form of a medical device. Further, a successful CHRIS evaluation does not address local biocompatibility endpoints,
25
- and therefore additional testing may be needed despite a positive outcome from CHRIS. Because there may be additional
26
- restrictions based on device specific guidance and code of federal regulation (CFR) listings, users should also review
27
- the relevant documents for their device type and color additive. </strong></p>
28
-
29
-
30
- <p> For details on how to use CHRIS, please read the <a href="{{url_for('static', filename='README.html')}}"> instructions.</a> and <a href="{{url_for('static', filename='COU.html')}}"> context of use (COU)</a>. Answers to frequently asked questions can be found <a href="{{url_for('static', filename='FAQ.html')}}"> here</a>.</p>
31
 
32
  <body>
33
 
 
11
 
12
  </head>
13
 
 
 
 
 
14
  <header>
15
+ <h1 style="text-align:center">CHRIS: <font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator</h1>
16
  </header>
17
 
18
+ <p> For details on how to use CHRIS, please read the <a href="{{url_for('static', filename='README.html')}}"> instructions.</a> and
19
+ <a href="{{url_for('static', filename='COU.html')}}"> context of use (COU)</a>, which includes limitations of use. Answers to frequently asked questions can be
20
+ found <a href="{{url_for('static', filename='FAQ.html')}}"> here</a>.</p>
 
 
 
 
 
 
21
 
22
  <body>
23
 
templates/report.html CHANGED
@@ -10,10 +10,6 @@
10
 
11
  </head>
12
 
13
- <img src="{{ url_for('static',filename='images/FDAlogo.png') }}" style="float: left;" height="100"/>
14
- <img src="{{ url_for('static',filename='images/FDAgraphic.png') }}" style="float: right;" height="100"/>
15
- <br clear="all" />
16
-
17
  <header>
18
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) Report </h1>
19
  </header>
 
10
 
11
  </head>
12
 
 
 
 
 
13
  <header>
14
  <h1 style="text-align:center"><font color="#0070C0">C</font>olor <font color="#0070C0">H</font>azard and <font color="#0070C0">RIS</font>k calculator (CHRIS) Report </h1>
15
  </header>