CAPTURE Dataset: GPDAS Prototype Data Acquisition
Collection Date: November 13, 2026
Environment: Benchtop colonoscopy model
System: GPDAS Instrumented Data Acquisition Prototype
Overview
This dataset contains five recording sessions collected using the GPDAS prototype, an instrumented colonoscopy testbed developed to capture synchronized video and operator actions.
The recordings combine intraluminal video, external camera views, sensor measurements, and timing information to support research in computer vision, sensor integration, and machine learning.
This collection represents an early stage of prototype development and is intended for research and development rather than clinical evaluation.
Recorded Data
| Source | Description |
|---|---|
| Ambu aBox | Internal endoscope video showing the simulated intraluminal environment |
| Basler0 | High-resolution external camera recording supporting observation of instrument activity |
| Logitech Cameras (2) | Supplementary external views of the instrument and procedure |
| Sensor and Timing Logs | Recorded sensor channels, frame timestamps, and synchronization information |
Sensor channels include demonstration measurements and should not be interpreted as fully calibrated operator-action data.
Recording Sessions
| Session | Description |
|---|---|
| Session 1 | Initial recording test, stops at simulated CECUM (FWD INSERTION ONLY)) |
| Session 2 | Short setup and synchronization test |
| Session 3 | Completed 24-minute recording with two verified optical synchronization markers |
| Session 4 | Completed procedure, LED End sequence missing due to drive space |
| Session 5 | Partial recording with recovered webcam footage |
Session 3 is the recommended starting point for researchers evaluating the available recordings and synchronization data. Its aBox video pairing was verified using both optical markers. Pairings for the remaining sessions are provisional.
Accessing the Recordings
Original camera recordings and associated metadata are preserved where available. Encoded MP4 examples are provided for easier inspection without requiring specialized raw-video processing tools.
Encoding Examples
- Basler MP4: Converted from the original raw Bayer image stream for conventional video playback.
- Logitech MP4: Recovered and corrected webcam recordings, with available frame timing information preserved separately.
- Ambu aBox: Internal endoscope recordings provided for comparison with external views.
Encoded video is intended for viewing and preliminary analysis. Researchers performing synchronization or frame-level processing should use the original recordings and associated timing metadata.
Working with Raw Video
Basler Camera
The Basler camera records uncompressed image frames in a raw Bayer format. Unlike an MP4, the raw file does not contain a conventional video header or playback timing information.
| File | Purpose |
|---|---|
basler0.raw |
Original raw image frames |
basler0_frames.csv |
Frame timing information |
Recording metadata (.json) |
Image dimensions, pixel format, and acquisition parameters |
The raw recording can be previewed or converted using FFmpeg, provided the correct image dimensions and pixel format are specified.
Example: Preview BayerRG8 video
ffplay -f rawvideo \
-pixel_format bayer_rggb8 \
-video_size WIDTHxHEIGHT \
-framerate FPS \
basler0.raw
Example: Convert to H.264 MP4
ffmpeg -f rawvideo \
-pixel_format bayer_rggb8 \
-video_size WIDTHxHEIGHT \
-framerate FPS \
-i basler0.raw \
-vf format=yuv420p \
-c:v libx264 -crf 18 \
basler0_converted.mp4
Replace WIDTH, HEIGHT, and FPS with the recording parameters. These examples assume tightly packed BayerRG8 frames without row padding; consult the recording metadata before processing.
The frame rate supplied during conversion determines playback speed. For synchronization and temporal analysis, use basler0_frames.csv rather than the timestamps generated by the converted MP4.
Logitech Cameras
Webcam recordings are provided as corrected MP4 files. Each reconstructed clip begins at its own first available frame.
- Sessions 2 and 3 use
camera*_frames.csv. - Sessions 1, 4, and 5 use
camera*_frame_map.csv.
The timing files support approximate alignment between camera streams. Cross-camera exposure timing is not guaranteed to be frame-accurate.
Endoscope Video and Sensor Data
The internal aBox recordings can be examined alongside external camera views and the logged sensor measurements.
Session 3 has a verified aBox video pairing using two optical synchronization markers. Other pairings remain provisional.
Data Quality and Limitations
This collection includes preliminary and partially recovered recordings. The following limitations apply:
- Sessions 1, 4, and 5 are incomplete and include recovered webcam footage.
- Partial Basler recordings may terminate with an incomplete image frame.
- Cross-camera synchronization is approximate.
- Frame timestamps do not necessarily represent exact exposure times.
- Sensor channels include demonstration measurements and are not fully calibrated.
- Calibrated insertion depth and clinical annotations are not included.
Researchers should review individual session metadata before performing synchronized or frame-level analysis.
Research Applications
The dataset supports preliminary investigation of multimodal data acquisition, video synchronization, optical motion tracking, sensor-video alignment, and machine learning workflows.
The recordings represent prototype research data and do not constitute a clinically validated or fully calibrated ground-truth dataset.
License and Usage
License information is pending. No public reuse license has been assigned.
- Downloads last month