| """CHEM-091 ECO PHYSICS nCLD 822 M dual-chamber chemiluminescence NO / NO2 / NOx analyzer: runtime (process model + session). | |
| What is simulated (SI; mole fractions in mol/mol, the GUI shows ppm = 1e-6 mol/mol with 0.1 ppm resolution): | |
| * inlet: the gas at the sample inlet is either room air (disconnected), a calibration cylinder from the template library | |
| (NIST SRM NO-in-N2 / O2-in-N2 / CO2-in-N2, connected by the agent) or a host-supplied sample (process truth, evaluator only); | |
| inlet pressure 600-1200 mbar absolute (brochure), sample flow 1.0 L/min at 101325 Pa through critical restrictors (flow ~ p_in); | |
| * sample path: optional external sample line (host: length, inner diameter, PTFE / stainless, temperature) with wall memory of NO and | |
| NO2 from NASA TM X-3229 (rise / decay time per wetted area); condensation in unheated paths (sample dew point above the path | |
| temperature, option "h" hot tubing absent) removes soluble species by Henry's-law equilibrium with the condensate (Sander 2023); | |
| dark gas-phase chemistry during the residence time (IUPAC rows 53-64: 2NO+O2, NO+O3, NO+NO3, NO2+O3, N2O3 / N2O4 / N2O5 | |
| formation and decomposition, N2O5 hydrolysis) integrated with backward Euler (N-atom conserving); | |
| * two chambers: NO channel = NO arriving; NOx channel = after the metal converter (NO2 efficiency = host parameter, default 1.0; | |
| other NOy by the qualitative reducibility classes of Dunlea et al. 2007: efficient -> 1, not efficient -> 0, partial -> explicit | |
| host assumption); O2-dependent NOx-channel bias from NASA TM X-3229 Table VII; | |
| * chemiluminescence: signal ~ NO x quench factor x flow, quench 1 / (1 + sum a_i x_i) with measured interference factors (CO2 0.60, | |
| H2O 2.1, O2 0) and, for Ar / H2 / N2O / NO2, a_i = q_i / q_N2 - 1 from the relative quenching efficiencies (upper bound: the sample | |
| is taken as the whole third-body bath); the instrument compensates the indicated inlet pressure (option "r" regulates it); | |
| * dynamics: internal lag 2.0 s at 1.0 L/min (brochure: lag < 3 s), mixing tau 0.35 s (0-90 % in 0.81 s, brochure < 1 s), filter | |
| tau = setting; zero noise 0.06 ppm (1 sigma) at the reference filter 1 s scaled with sqrt(1 s / tau) (brochure: 0.06 ppm, LDL | |
| 0.12 ppm "depending on filter setting"); warm-up of converter and cooled PMT (time constants and set points are assumptions); | |
| * four freely selectable full-scale ranges 5-5000 ppm (over-range is flagged), zero / span calibration (asynchronous averaging), | |
| 1 Hz time series of NO, NO2 = NOx - NO and NOx while measuring. | |
| Not simulated: V1 / V2 calibration valves, the analog output box, data interfaces, fault probabilities, PMT gain drift. | |
| Matterix bridge: RECIPES module="asset_runtime.eco_ncld822", cls="EcoNcld822Runtime", instance_in_ctor=True, advance=STEP_DT.""" | |
| from __future__ import annotations | |
| import copy | |
| import hashlib | |
| import json | |
| import math | |
| import random | |
| from collections import deque | |
| from pathlib import Path | |
| import numpy as np | |
| from asset_runtime import _annotations as _ann | |
| PROTOCOL_VERSION = '1.0.0' | |
| ASSET_TYPE_ID = 'instrument.eco_ncld822' | |
| CATALOG_ID = 'CHEM-091' | |
| ROOT = Path(__file__).resolve().parents[2] | |
| TEMPLATE_REL = 'assets/reactions/eco_ncld822_nox_library/templates.json' | |
| R_GAS, K_B = 8.314462618, 1.380649e-23 | |
| P_REF, T_REF = 101325.0, 298.15 | |
| PPM = 1e-6 | |
| Q_NOM = 1.0e-3 / 60.0 # 1.0 L/min (brochure), taken at 101325 Pa / 298.15 K (assumption: no reference state given) | |
| P_MIN, P_MAX = 60000.0, 120000.0 # 600-1200 mbar abs (brochure) | |
| FS_MIN, FS_MAX = 5 * PPM, 5000 * PPM # four freely selectable ranges 5-5000 ppm (brochure) | |
| LAG_INT_S = 2.0 # internal transport lag at Q_NOM (assumption within the brochure's < 3 s) | |
| TAU_MIX_S = 0.35 # chamber / path mixing (assumption; 0-90 % = 2.303 tau = 0.81 s < 1 s brochure) | |
| NOISE0 = 0.06 * PPM # zero noise 1 sigma (brochure, "depending on filter setting") | |
| FILTER_REF_S = 1.0 # filter setting at which NOISE0 holds (assumption) | |
| TAU_P_SENSOR_S = 0.5 # inlet pressure sensor lag used by the pressure compensation (assumption) | |
| T_CONV_SET, TAU_CONV_S = 648.15, 150.0 # metal converter set point / warm-up tau (assumptions: brochure gives neither) | |
| T_PMT_SET, TAU_PMT_S = 268.15, 90.0 # cooled PMT set point / tau (assumptions) | |
| TAU_O3_S = 20.0 # ozone generator ramp (assumption) | |
| T_HOT_S = 453.15 # option "h" hot tubing temperature (assumption) | |
| DIGITS = 1e-7 # 0.1 ppm display resolution (GUI screenshot) | |
| SERIES_DT = 1.0 # 1 Hz series while measuring (assumption; brochure: second-level signals) | |
| SERIES_MAX = 7200 | |
| DEFAULT_FS = [5e-6, 5e-5, 5e-4, 5e-3] | |
| STANDARD_EXTRA = {'zero_nitrogen': dict(nitrogen=1.0), 'zero_air': dict(nitrogen=0.7808, oxygen=0.2095, argon=0.0093, carbon_dioxide=0.0004)} | |
| CHEM = ['nitric_oxide', 'nitrogen_dioxide', 'ozone', 'nitrate_radical', 'dinitrogen_pentoxide', 'nitric_acid', 'dinitrogen_trioxide', | |
| 'dinitrogen_tetroxide', 'oxygen', 'water'] | |
| N_ATOMS = {'nitric_oxide': 1, 'nitrogen_dioxide': 1, 'nitrate_radical': 1, 'dinitrogen_pentoxide': 2, 'nitric_acid': 1, 'dinitrogen_trioxide': 2, | |
| 'dinitrogen_tetroxide': 2, 'nitrous_acid': 1, 'ammonia': 1, 'peroxyacetyl_nitrate': 1, 'peroxynitric_acid': 1, 'methyl_nitrate': 1, | |
| 'ethyl_nitrate': 1, 'n_propyl_nitrate': 1, 'n_butyl_nitrate': 1, 'nitrous_oxide': 2, 'hydrogen_cyanide': 1, 'acetonitrile': 1, | |
| 'nitromethane': 1, 'particulate_nitrate': 1} | |
| NO_PER_MOLECULE = {'nitrogen_dioxide': 1, 'nitrate_radical': 1, 'dinitrogen_pentoxide': 2, 'nitric_acid': 1, 'dinitrogen_trioxide': 2, | |
| 'dinitrogen_tetroxide': 2, 'methyl_nitrate': 1, 'ethyl_nitrate': 1, 'n_propyl_nitrate': 1, 'n_butyl_nitrate': 1, | |
| 'nitrous_acid': 1, 'ammonia': 1, 'peroxyacetyl_nitrate': 1, 'peroxynitric_acid': 1, 'nitrous_oxide': 2, 'hydrogen_cyanide': 1, | |
| 'acetonitrile': 1, 'nitromethane': 1, 'particulate_nitrate': 1} | |
| CLASS_ALIAS = {'peroxyacetyl_nitrate': 'peroxy_nitrates_ro2no2', 'dinitrogen_trioxide': 'nitrogen_dioxide', 'dinitrogen_tetroxide': 'nitrogen_dioxide'} | |
| class ActionRejected(ValueError): | |
| """Precondition or argument failure; the runtime state is unchanged. Messages never disclose sample composition.""" | |
| # ------------------------------------------------------------------------------------------------------------------- template library | |
| _LIB = None | |
| def library(): | |
| global _LIB | |
| if _LIB is None: | |
| path = ROOT / TEMPLATE_REL | |
| raw = path.read_bytes() | |
| data = json.loads(raw) | |
| _LIB = dict(sha256=hashlib.sha256(raw).hexdigest(), by_id={r['record_id']: r for r in data['templates']}, data=data) | |
| return _LIB | |
| def _rec(rid): | |
| return library()['by_id'][rid] | |
| def _v(rid, key): | |
| return _rec(rid)['values'][key]['value'] | |
| def certificate_validity(standard_id): | |
| c = library()['by_id'][standard_id]['conditions'] | |
| return c.get('validity') | |
| def certificates(): | |
| """{standard_id: (species, certificate value, U95, value kind)} of the connectable SRM cylinders. Individually certified lots have no | |
| public value: the simulated cylinder's certificate is taken at the nominal fraction (assumption, flagged in the result).""" | |
| out = {} | |
| for rid, r in library()['by_id'].items(): | |
| if r['family'] != 'calibration_standard': | |
| continue | |
| v = r['values'] | |
| cert = v.get('certified_mole_fraction') or v.get('example_certified_mole_fraction') | |
| x = (cert or v['nominal_mole_fraction'])['value'] | |
| out[rid] = (r['substance_id'], float(x), float(v['expanded_uncertainty_U95']['value']), | |
| r['conditions'].get('value_kind') if cert else 'nominal_used_individually_certified') | |
| return out | |
| def standards(): | |
| """Certificate compositions of the connectable cylinders (balance N2) + two built-in zero gases.""" | |
| out = {rid: {sp: x, 'nitrogen': 1.0 - x} for rid, (sp, x, u, kind) in certificates().items()} | |
| out.update(copy.deepcopy(STANDARD_EXTRA)) | |
| return out | |
| def cylinder_truth(instance_id, standard_id): | |
| """True mole fraction of the simulated cylinder: certificate value x (1 + d), d ~ N(0, (U95 / 2) / x) drawn deterministically from the | |
| instance and standard ids (the cylinder content is unknown within its certified uncertainty).""" | |
| sp, x, u, kind = certificates()[standard_id] | |
| h = hashlib.sha256(f'{instance_id}|{standard_id}'.encode()).digest() | |
| u1, u2 = (int.from_bytes(h[:8], 'big') + 0.5) / 2 ** 64, (int.from_bytes(h[8:16], 'big') + 0.5) / 2 ** 64 | |
| z = math.sqrt(-2.0 * math.log(u1)) * math.cos(2.0 * math.pi * u2) | |
| return sp, x * (1.0 + z * (u / 2.0) / x) | |
| def quench_coefficients(phi=0.0): | |
| """a_i in S/S0 = 1 / (1 + sum a_i x_i), relative to a nitrogen bath. CO2 / H2O / O2: measured interference factors (NASA TM X-3229). | |
| Ar / H2 / N2O / NO2: relative quenching efficiencies (Clyne et al. via the same report) scaled by phi = the fraction of the chamber's third | |
| bodies contributed by the sample; phi is an explicit host assumption (default 0: the measured O2 factor ~0 contradicts phi = 1, which would | |
| give O2 +11 % over N2).""" | |
| a = {} | |
| for rid in ('quench_if_carbon_dioxide', 'quench_if_water', 'quench_if_oxygen'): | |
| a[_rec(rid)['substance_id']] = 0.0 - float(_v(rid, 'dlnS_dx')) | |
| q_n2 = float(_v('quench_rel_nitrogen', 'relative_quenching_efficiency_k3M_over_k3O2')) | |
| for rid in ('quench_rel_argon', 'quench_rel_hydrogen', 'quench_rel_nitrous_oxide', 'quench_rel_nitrogen_dioxide'): | |
| sid = _rec(rid)['substance_id'] | |
| a.setdefault(sid, phi * (float(_v(rid, 'relative_quenching_efficiency_k3M_over_k3O2')) / q_n2 - 1.0)) | |
| return a | |
| def converter_class(sid): | |
| rid = 'converter_' + CLASS_ALIAS.get(sid, sid) | |
| r = library()['by_id'].get(rid) | |
| return r['values']['reducibility_class']['value'] if r else None | |
| def henry_constant(sid, T): | |
| r = library()['by_id'].get('henry_' + sid) | |
| if r is None: | |
| return None | |
| h = float(r['values']['Hscp_298']['value']) | |
| b = r['values'].get('temperature_coefficient_B', {}).get('value') | |
| return h * math.exp(float(b) * (1.0 / T - 1.0 / 298.15)) if b is not None else h | |
| def adsorption_tau(sid, material, area_m2): | |
| """(tau_rise_s, tau_decay_s) of the line wall memory: time per wetted area x area / ln(1000) (0.1 % criterion of the report).""" | |
| r = library()['by_id'].get(f"line_adsorption_{sid}_{ {'ptfe': 'teflon'}.get(material, material)}") | |
| if r is None or area_m2 <= 0: | |
| return 0.0, 0.0 | |
| rise = r['values']['rise_time_per_area']['value'] | |
| decay = r['values']['decay_time_per_area']['value'] | |
| rise = decay if rise is None else rise | |
| return float(rise) * area_m2 / math.log(1000.0), float(decay) * area_m2 / math.log(1000.0) | |
| def o2_converter_bias(x_o2): | |
| v = _rec('converter_o2_effect_stainless_800c')['values'] | |
| xs, es = v['oxygen_mole_fraction']['value'], v['nox_mode_relative_error']['value'] | |
| if x_o2 <= xs[0]: | |
| return es[0] | |
| for i in range(1, len(xs)): | |
| if x_o2 <= xs[i]: | |
| return es[i - 1] + (es[i] - es[i - 1]) * (x_o2 - xs[i - 1]) / (xs[i] - xs[i - 1]) | |
| return es[-1] | |
| def p_sat_water(T): | |
| """Saturation vapour pressure of water over liquid (Magnus-Tetens / Alduchov-Eskridge 1996 form; engineering formula).""" | |
| tc = T - 273.15 | |
| return 610.94 * math.exp(17.625 * tc / (tc + 243.04)) | |
| # ------------------------------------------------------------------------------------------------------------------- dark chemistry | |
| def _arr(v, T, key='A', ekey='E_over_R'): | |
| return float(v[key]['value']) * math.exp(-float(v[ekey]['value']) / T) | |
| def _clamp(T, rng): | |
| return min(max(T, rng[0]), rng[1]) if rng else T | |
| def _troe(v, T, n_m, clamped=None, row=None): | |
| """IUPAC Troe falloff; k0 and kinf are each evaluated with T clamped into their own evaluated range.""" | |
| k0p, kip = v['k0_prefactor'], v['kinf_prefactor'] | |
| T0, Ti = _clamp(T, k0p.get('T_range_K')), _clamp(T, kip.get('T_range_K')) | |
| if clamped is not None and (T0 != T or Ti != T): | |
| clamped.append(row) | |
| k0 = float(k0p['value']) * (T0 / 300.0) ** float(k0p['T_exponent_n']) | |
| if k0p.get('exp_minus_E_over_RT_K'): | |
| k0 *= math.exp(-float(k0p['exp_minus_E_over_RT_K']) / T0) | |
| ki = float(kip['value']) * (Ti / 300.0) ** float(kip['T_exponent_n']) | |
| if kip.get('exp_minus_E_over_RT_K'): | |
| ki *= math.exp(-float(kip['exp_minus_E_over_RT_K']) / Ti) | |
| k0m = k0 * n_m | |
| fc = float(v['Fc']['value']) | |
| nn = 0.75 - 1.27 * math.log10(fc) | |
| logf = math.log10(fc) / (1.0 + (math.log10(k0m / ki) / nn) ** 2) | |
| return k0m * ki / (k0m + ki) * 10 ** logf | |
| def t_range(v): | |
| return v['T_range']['value'] if isinstance(v.get('T_range'), dict) else None | |
| def rate_constants(T, n_m, clamped=None): | |
| """SI rate coefficients (molecule, m^3, s) of IUPAC rows 53-64 at T and bath number density n_m; T is clamped into each record's | |
| evaluated temperature range (the clamped rows are appended to `clamped`).""" | |
| k = {} | |
| T0 = T | |
| for i in range(53, 65): | |
| r = _rec(f'kinetics_iupac_{i}') | |
| law, v = r['conditions']['rate_law'], r['values'] | |
| if law.startswith('troe'): | |
| k[i] = _troe(v, T0, n_m, clamped, i) | |
| continue | |
| T = _clamp(T0, t_range(v)) | |
| if clamped is not None and T != T0: | |
| clamped.append(i) | |
| if law in ('termolecular_arrhenius', 'bimolecular_arrhenius'): | |
| k[i] = _arr(v, T) | |
| else: | |
| k[i] = float(v['k']['value']) | |
| return k | |
| def _rates(n, k): | |
| no, no2, o3, no3, n2o5, hno3, n2o3, n2o4, o2, h2o = n | |
| return np.array([k[53] * no * no * o2, k[54] * no * o3, k[55] * no * no2, k[56] * n2o3, k[57] * no * no3, k[58] * no2 * o3, | |
| k[59] * no2 * no2, k[60] * n2o4, k[61] * no2 * no3, k[62] * n2o5, k[63] * n2o5 * h2o, k[64] * n2o5 * h2o * h2o]) | |
| # stoichiometry: rows = species (CHEM order), columns = reactions 53..64 | |
| _S = np.zeros((10, 12)) | |
| _REACTIONS = [{0: -2, 8: -1, 1: 2}, {0: -1, 2: -1, 1: 1, 8: 1}, {0: -1, 1: -1, 6: 1}, {6: -1, 0: 1, 1: 1}, {0: -1, 3: -1, 1: 2}, | |
| {1: -1, 2: -1, 3: 1, 8: 1}, {1: -2, 7: 1}, {7: -1, 1: 2}, {1: -1, 3: -1, 4: 1}, {4: -1, 1: 1, 3: 1}, {4: -1, 9: -1, 5: 2}, | |
| {4: -1, 9: -1, 5: 2}] | |
| for _j, _nu in enumerate(_REACTIONS): | |
| for _i, _c in _nu.items(): | |
| _S[_i, _j] = _c | |
| def integrate_dark_chemistry(x, T, p, t_res, clamped=None): | |
| """Mole fractions of the CHEM species after t_res seconds at (T, p): backward Euler + Newton on number densities, geometric substeps. | |
| Linear invariants (N atoms, O atoms) are preserved by construction up to the Newton tolerance.""" | |
| n_m = p / (K_B * T) | |
| n0 = np.array([x.get(s, 0.0) for s in CHEM]) * n_m | |
| if t_res <= 0 or (n0[0] <= 0 and n0[1] <= 0 and n0[3] <= 0 and n0[4] <= 0 and n0[6] <= 0 and n0[7] <= 0): | |
| return {s: float(v / n_m) for s, v in zip(CHEM, n0)} | |
| k = rate_constants(T, n_m, clamped) | |
| n = n0.copy() | |
| t, h = 0.0, min(1e-7, t_res) | |
| while t < t_res - 1e-15: | |
| h = min(h, t_res - t) | |
| y = n.copy() | |
| for _ in range(30): | |
| f = y - n - h * (_S @ _rates(y, k)) | |
| J = np.eye(10) | |
| for i in range(10): # finite-difference Jacobian of the rates (10 x 10) | |
| d = max(abs(y[i]) * 1e-7, 1e-30 * n_m + 1.0) | |
| yp = y.copy() | |
| yp[i] += d | |
| J[:, i] -= h * (_S @ (_rates(yp, k) - _rates(y, k))) / d | |
| dy = np.linalg.solve(J, -f) | |
| y = y + dy | |
| if np.all(np.abs(dy) <= 1e-10 * np.maximum(np.abs(y), 1e-12 * n_m)): | |
| break | |
| y = np.where(y < 0, np.where(y > -1e-12 * n_m, 0.0, y), y) | |
| n = y | |
| t += h | |
| h *= 1.6 | |
| return {s: float(v / n_m) for s, v in zip(CHEM, n)} | |
| def n_atoms(x): | |
| return sum(N_ATOMS.get(s, 0) * v for s, v in x.items()) | |
| # ------------------------------------------------------------------------------------------------------------------- schemas | |
| def number(unit, lo=None, hi=None): | |
| d = {'type': 'number', 'x-unit': unit} | |
| if lo is not None: | |
| d['minimum'] = lo | |
| if hi is not None: | |
| d['maximum'] = hi | |
| return d | |
| def obj(props, required=None): | |
| return dict(type='object', properties=props, required=list(props) if required is None else required, additionalProperties=False) | |
| def nullable(s): | |
| return {'anyOf': [s, {'type': 'null'}]} | |
| BOOL, TEXT = {'type': 'boolean'}, {'type': 'string'} | |
| FRACTION = number('mol/mol', 0.0, 1.0) | |
| FS_ARRAY = {'type': 'array', 'x-unit': 'mol/mol', 'items': number('mol/mol', FS_MIN, FS_MAX), 'minItems': 4, 'maxItems': 4} | |
| STATUS = {'type': 'string', 'enum': ['off', 'warming', 'ready']} | |
| MEAS = {'type': 'string', 'enum': ['idle', 'measuring']} | |
| CALP = {'type': 'string', 'enum': ['idle', 'zeroing', 'spanning']} | |
| FLAGS = {'type': 'array', 'items': {'type': 'string', 'enum': ['warming_up', 'over_range', 'inlet_pressure_out_of_range', 'ambient_out_of_range', | |
| 'no_ozone', 'calibration_running']}} | |
| MATERIAL = {'type': 'string', 'enum': ['ptfe', 'stainless']} | |
| COMPOSITION = {'type': 'object', 'additionalProperties': number('mol/mol', 0.0, 1.0)} | |
| POINT = obj({'time_s': number('s', 0), 'no_mol_mol': nullable(number('mol/mol')), 'no2_mol_mol': nullable(number('mol/mol')), | |
| 'nox_mol_mol': nullable(number('mol/mol'))}) | |
| PHASE_OP = {'type': 'string', 'enum': ['running', 'succeeded', 'cancelled', 'failed']} | |
| CAL_RESULT = obj({'phase': CALP, 'operation_id': TEXT}) | |
| MEASURE = obj({'status': STATUS, 'measurement': MEAS, 'quality': {'type': 'string', 'enum': ['valid', 'invalid', 'off']}, | |
| 'no_mol_mol': nullable(number('mol/mol')), 'no2_mol_mol': nullable(number('mol/mol')), 'nox_mol_mol': nullable(number('mol/mol')), | |
| 'no_ppm': nullable(number('ppm')), 'no2_ppm': nullable(number('ppm')), 'nox_ppm': nullable(number('ppm')), | |
| 'range_index': {'type': 'integer', 'minimum': 0, 'maximum': 3}, 'full_scale_mol_mol': number('mol/mol', FS_MIN, FS_MAX), | |
| 'flags': FLAGS, 'series': {'type': 'array', 'items': POINT}, 'template_source': TEXT, 'library_sha256': TEXT, | |
| 'template_records': {'type': 'array', 'items': TEXT}, | |
| 'evidence': {'type': 'string', 'const': 'simulated_from_cited_response_parameters'}}) | |
| STANDARD_IDS = sorted(standards()) | |
| MODEL_RECORDS = sorted(library()['by_id']) # every record family is used by the response model (static: no per-sample leak) | |
| PLANT = obj({'hot_tubing_installed': BOOL, 'pressure_regulation_installed': BOOL, 'line_length_m': number('m', 0.0, 50.0), | |
| 'line_inner_diameter_m': number('m', 0.001, 0.02), 'line_material': MATERIAL, 'line_temperature_k': number('K', 253.15, 473.15), | |
| 'ambient_temperature_k': number('K', 253.15, 333.15), 'ambient_pressure_pa': number('Pa', 60000.0, 120000.0), | |
| 'ambient_relative_humidity': number('1', 0.0, 1.0), 'converter_no2_efficiency': number('1', 0.0, 1.0), | |
| 'partial_conversion_fraction': nullable(number('1', 0.0, 1.0)), 'quench_relative_scale': number('1', 0.0, 1.0), 'noise_enabled': BOOL, 'seed': {'type': 'integer', 'minimum': 0, 'x-unit': '1'}}, | |
| required=[]) | |
| PLANT_EXAMPLE = dict(hot_tubing_installed=False, pressure_regulation_installed=False, line_length_m=2.0, line_inner_diameter_m=0.004, | |
| line_material='ptfe', line_temperature_k=298.15, ambient_temperature_k=298.15, ambient_pressure_pa=101325.0, | |
| ambient_relative_humidity=0.5, converter_no2_efficiency=1.0, partial_conversion_fraction=None, quench_relative_scale=0.0, | |
| noise_enabled=True, seed=0) | |
| INLET = obj({'composition': COMPOSITION, 'pressure_pa': number('Pa', 10000.0, 300000.0), 'temperature_k': number('K', 253.15, 773.15), 'label': TEXT}) | |
| INLET_EXAMPLE = dict(composition={'nitric_oxide': 250e-6, 'nitrogen_dioxide': 20e-6, 'oxygen': 0.06, 'carbon_dioxide': 0.08, 'water': 0.012}, | |
| pressure_pa=101325.0, temperature_k=298.15, label='scr_outlet') | |
| # capability: input, output, components, meaning, kind, visibility, example, state effects | |
| CAPS = { | |
| 'power.set': (obj({'enabled': BOOL}), obj({'enabled': BOOL}), ['switch_power', 'power_inlet', 'touchscreen'], | |
| '整机通断;断电停止测量、取消校准,转换器与PMT回到环境温度', 'device_command', 'agent', {'enabled': True}, | |
| ['power.enabled', 'status.phase', 'measurement.phase', 'calibration.phase']), | |
| 'power.toggle': (obj({}), obj({'enabled': BOOL}), ['switch_power'], '电源翘板按一次切换通断(Matterix按键)', 'device_command', 'agent', {}, | |
| ['power.enabled', 'status.phase']), | |
| 'range.configure': (obj({'full_scales_mol_mol': FS_ARRAY}), obj({'full_scales_mol_mol': FS_ARRAY}), ['touchscreen', 'display_digits'], | |
| '四个可自由选择的满量程(5–5000 ppm,升序)', 'configuration', 'agent', {'full_scales_mol_mol': DEFAULT_FS}, ['range.full_scales_mol_mol']), | |
| 'range.select': (obj({'index': {'type': 'integer', 'minimum': 0, 'maximum': 3, 'x-unit': '1'}}), obj({'index': {'type': 'integer', 'minimum': 0, 'maximum': 3, 'x-unit': '1'}}), | |
| ['touchscreen', 'display_digits'], '选择当前量程(超量程只标记,不自动切换)', 'configuration', 'agent', {'index': 2}, | |
| ['range.index', 'range.active_full_scale_mol_mol']), | |
| 'filter.set': (obj({'time_constant_s': number('s', 0.0, 60.0)}), obj({'time_constant_s': number('s', 0.0, 60.0)}), ['touchscreen', 'detector_block'], | |
| '输出一阶滤波时间常数;噪声与最低可检浓度随之变化', 'configuration', 'agent', {'time_constant_s': 1.0}, ['filter.time_constant_s']), | |
| 'measurement.start': (obj({}), obj({'phase': MEAS}), ['softkey_measure', 'detector_block', 'converter', 'vacuum_pump', 'display_digits'], | |
| '开始连续测量并以1 Hz记录NO/NO2/NOx时间序列(需预热完成)', 'device_command', 'agent', {}, ['measurement.phase', 'measurement.points']), | |
| 'measurement.stop': (obj({}), obj({'phase': MEAS}), ['softkey_measure', 'display_digits'], '停止连续测量(保留已记录序列)', 'device_command', 'agent', {}, | |
| ['measurement.phase']), | |
| 'measurement.toggle': (obj({}), obj({'phase': MEAS}), ['softkey_measure'], '测量触摸键:开始/停止切换', 'device_command', 'agent', {}, ['measurement.phase']), | |
| 'calibration.configure': (obj({'span_no_mol_mol': number('mol/mol', FS_MIN * 0.2, FS_MAX), 'span_nox_mol_mol': number('mol/mol', FS_MIN * 0.2, FS_MAX), | |
| 'window_s': number('s', 5.0, 600.0)}), | |
| obj({'span_no_mol_mol': number('mol/mol'), 'span_nox_mol_mol': number('mol/mol'), 'window_s': number('s')}), | |
| ['touchscreen'], '输入跨度气证书浓度(NO与NOx通道)与平均窗口', 'configuration', 'agent', | |
| {'span_no_mol_mol': 2954e-6, 'span_nox_mol_mol': 2954e-6, 'window_s': 30.0}, | |
| ['calibration.span_no_mol_mol', 'calibration.span_nox_mol_mol', 'calibration.window_s']), | |
| 'calibration.zero': (obj({}), CAL_RESULT, ['softkey_zero', 'detector_block', 'converter'], '零点校准:在入口零气下按窗口平均两通道读数并写入零点偏置(异步)', | |
| 'device_command', 'agent', {}, ['calibration.phase', 'calibration.zero_offset_no_mol_mol', 'calibration.zero_offset_nox_mol_mol']), | |
| 'calibration.span': (obj({}), CAL_RESULT, ['softkey_span', 'detector_block', 'converter'], '跨度校准:在入口跨度气下按窗口平均并按证书浓度写入增益(异步)', | |
| 'device_command', 'agent', {}, ['calibration.phase', 'calibration.gain_no', 'calibration.gain_nox']), | |
| 'calibration.cancel': (obj({}), obj({'phase': CALP}), ['touchscreen'], '取消正在进行的校准,保留原校准系数', 'device_command', 'agent', {}, ['calibration.phase']), | |
| 'inlet.connect_standard': (obj({'standard_id': {'type': 'string', 'enum': STANDARD_IDS}}), | |
| obj({'standard_id': TEXT, 'source': TEXT, 'certificate_mole_fraction': nullable(number('mol/mol', 0, 1)), | |
| 'expanded_uncertainty_U95': nullable(number('mol/mol', 0, 1)), 'value_kind': TEXT, | |
| 'certificate_validity': nullable(TEXT), 'certificate_expired': BOOL}), | |
| ['sample_inlet', 'inlet_filter'], '把标准气瓶(模板库NIST SRM或零气)经减压阀接到样气入口(101325 Pa,有放空)', 'device_command', | |
| 'agent', {'standard_id': 'srm_2631a'}, ['inlet.source']), | |
| 'inlet.disconnect': (obj({}), obj({'source': TEXT}), ['sample_inlet'], '断开入口:泵抽入室内空气', 'device_command', 'agent', {}, ['inlet.source']), | |
| 'inlet.connect': (INLET, obj({'source': TEXT}), ['sample_inlet', 'inlet_filter'], '宿主接入样品气(组成、绝压、温度为过程真值)', 'configuration', | |
| 'evaluator', INLET_EXAMPLE, ['inlet.source', 'inlet.composition', 'inlet.pressure_pa', 'inlet.temperature_k']), | |
| 'plant.configure': (PLANT, obj({'configured': BOOL}), ['housing', 'converter', 'sample_inlet'], | |
| '宿主配置已装选件(h热管路、r压力调节)、外部样气管、环境与转换器参数', 'configuration', 'evaluator', PLANT_EXAMPLE, | |
| ['line.condensed_water_fraction']), | |
| 'nox.measure': (obj({'max_points': {'type': 'integer', 'minimum': 0, 'maximum': SERIES_MAX, 'x-unit': '1'}}, required=[]), MEASURE, | |
| ['display_digits', 'detector_block', 'converter'], '读取当前NO/NO2/NOx显示值、量程、状态标志与时间序列(纯读取,不推进时钟)', 'query', | |
| 'agent', {'max_points': 60}, []), | |
| } | |
| UNSUPPORTED = { | |
| 'calibration.valve_select': ('V1/V2校准电磁阀选件(单阀或零/跨度双阀、2–3 bar加压校准)未建模', ['sample_inlet']), | |
| 'analog_output.configure': ('外接模拟输出盒(0–10 V,4–20 mA)未建模', ['power_inlet']), | |
| 'data.export': ('USB/LAN/WLAN/蓝牙/HDMI数据通信与导出未实现', ['usb_ports', 'lan_port', 'hdmi_port']), | |
| } | |
| # slot: schema, unit, origin, visibility, components, write (capability, argument), meaning | |
| SLOTS = { | |
| 'power.enabled': (BOOL, '1', 'setpoint', 'agent', ['switch_power'], ('power.set', 'enabled'), '整机电源设定'), | |
| 'status.phase': (STATUS, '1', 'derived', 'agent', ['touchscreen'], None, '预热状态:off/warming/ready(转换器±10 K、PMT±1 K内为ready)'), | |
| 'status.flags': (FLAGS, '1', 'derived', 'agent', ['touchscreen'], None, 'GUI状态标志(预热、超量程、入口压力越界、环境越界、无臭氧、校准中、部分转换按假设)'), | |
| 'converter.temperature_k': (number('K', 0), 'K', 'process_truth', 'evaluator', ['converter'], None, '金属转换器温度真值(设定点648 K为假设)'), | |
| 'pmt.temperature_k': (number('K', 0), 'K', 'process_truth', 'evaluator', ['detector_block'], None, '冷却PMT温度真值(设定点268 K为假设)'), | |
| 'ozone.generator_fraction': (number('1', 0, 1), '1', 'process_truth', 'evaluator', ['ozone_generator'], None, '臭氧发生器相对输出(0–1)'), | |
| 'range.full_scales_mol_mol': (FS_ARRAY, 'mol/mol', 'setpoint', 'agent', ['touchscreen'], ('range.configure', 'full_scales_mol_mol'), '四个满量程'), | |
| 'range.index': ({'type': 'integer', 'minimum': 0, 'maximum': 3, 'x-unit': '1'}, '1', 'setpoint', 'agent', ['touchscreen'], ('range.select', 'index'), '当前量程序号'), | |
| 'range.active_full_scale_mol_mol': (number('mol/mol', FS_MIN, FS_MAX), 'mol/mol', 'derived', 'agent', ['display_digits'], None, '当前量程满量程'), | |
| 'filter.time_constant_s': (number('s', 0.0, 60.0), 's', 'setpoint', 'agent', ['touchscreen'], ('filter.set', 'time_constant_s'), '输出滤波时间常数'), | |
| 'measurement.phase': (MEAS, '1', 'derived', 'agent', ['display_digits'], None, '连续测量状态'), | |
| 'measurement.points': ({'type': 'integer', 'minimum': 0}, '1', 'derived', 'agent', ['display_digits'], None, '已记录的1 Hz序列点数'), | |
| 'calibration.phase': (CALP, '1', 'derived', 'agent', ['touchscreen'], None, '校准状态'), | |
| 'calibration.span_no_mol_mol': (number('mol/mol'), 'mol/mol', 'setpoint', 'agent', ['touchscreen'], None, 'NO通道跨度气证书浓度'), | |
| 'calibration.span_nox_mol_mol': (number('mol/mol'), 'mol/mol', 'setpoint', 'agent', ['touchscreen'], None, 'NOx通道跨度气证书浓度'), | |
| 'calibration.window_s': (number('s', 5.0, 600.0), 's', 'setpoint', 'agent', ['touchscreen'], None, '校准平均窗口'), | |
| 'calibration.zero_offset_no_mol_mol': (number('mol/mol'), 'mol/mol', 'derived', 'agent', ['detector_block'], None, 'NO通道零点偏置(归一化信号单位)'), | |
| 'calibration.zero_offset_nox_mol_mol': (number('mol/mol'), 'mol/mol', 'derived', 'agent', ['detector_block'], None, 'NOx通道零点偏置'), | |
| 'calibration.gain_no': (number('1', 0), '1', 'derived', 'agent', ['detector_block'], None, 'NO通道增益(出厂1.0)'), | |
| 'calibration.gain_nox': (number('1', 0), '1', 'derived', 'agent', ['detector_block'], None, 'NOx通道增益(出厂1.0)'), | |
| 'inlet.source': (TEXT, '1', 'derived', 'agent', ['sample_inlet'], None, '入口接的是什么:ambient_air / standard:<id> / sample(宿主样品,标签只对evaluator可见)'), | |
| 'inlet.indicated_pressure_pa': (number('Pa', 0), 'Pa', 'derived', 'agent', ['sample_inlet', 'touchscreen'], None, '入口绝压显示(一阶0.5 s,1 Pa分辨率)'), | |
| 'flow.indicated_sample_m3_s': (number('m^3/s', 0), 'm^3/s', 'derived', 'agent', ['flow_restrictors', 'touchscreen'], None, '样气体积流量显示'), | |
| 'inlet.composition': (COMPOSITION, 'mol/mol', 'process_truth', 'evaluator', ['sample_inlet'], None, '入口气体组成真值(含平衡N2)'), | |
| 'inlet.pressure_pa': (number('Pa', 0), 'Pa', 'process_truth', 'evaluator', ['sample_inlet'], None, '入口绝压真值'), | |
| 'inlet.temperature_k': (number('K', 0), 'K', 'process_truth', 'evaluator', ['sample_inlet'], None, '入口气体温度真值'), | |
| 'chamber.no_mol_mol': (number('mol/mol', 0), 'mol/mol', 'process_truth', 'evaluator', ['detector_block'], None, 'NO反应室的等效NO摩尔分数真值(传输、混合后)'), | |
| 'chamber.nox_mol_mol': (number('mol/mol', 0), 'mol/mol', 'process_truth', 'evaluator', ['detector_block', 'converter'], None, 'NOx反应室(转换后)的等效NO真值'), | |
| 'inlet.label': (nullable(TEXT), '1', 'process_truth', 'evaluator', ['sample_inlet'], None, '宿主样品的标签(agent只看到sample)'), | |
| 'converter.partial_assumption_applied': (BOOL, '1', 'process_truth', 'evaluator', ['converter'], None, | |
| '到达气中含“部分还原”类物种且使用了缺省0.5假设(未由宿主给定)'), | |
| 'line.condensed_water_fraction': (number('mol/mol', 0, 1), 'mol/mol', 'process_truth', 'evaluator', ['sample_inlet'], None, '未加热路径中冷凝的水(每摩尔样气)'), | |
| 'display.no_mol_mol': (nullable(number('mol/mol')), 'mol/mol', 'derived', 'agent', ['display_digits'], None, '屏幕NO读数(0.1 ppm分辨率;关机为null)'), | |
| 'display.no2_mol_mol': (nullable(number('mol/mol')), 'mol/mol', 'derived', 'agent', ['display_digits'], None, '屏幕NO2读数=NOx−NO'), | |
| 'display.nox_mol_mol': (nullable(number('mol/mol')), 'mol/mol', 'derived', 'agent', ['display_digits'], None, '屏幕NOx读数'), | |
| } | |
| SCENE_MIRRORED = ('display.no_mol_mol', 'display.no2_mol_mol', 'display.nox_mol_mol', 'power.enabled', 'status.phase', 'measurement.phase', 'calibration.phase') | |
| REFERENCE_ZH = { | |
| 'inlet.indicated_pressure_pa': '绝压(厂家进口600–1200 mbar abs.)。', | |
| 'inlet.pressure_pa': '绝压。', | |
| 'flow.indicated_sample_m3_s': '气体体积流量,按101325 Pa、298.15 K参考状态(假设;厂家1.0 l/min未注明参考状态);临界限流下与入口绝压成正比。', | |
| } | |
| for _k in ('range.full_scales_mol_mol', 'range.active_full_scale_mol_mol', 'calibration.span_no_mol_mol', 'calibration.span_nox_mol_mol', | |
| 'display.no_mol_mol', 'display.no2_mol_mol', 'display.nox_mol_mol', 'chamber.no_mol_mol', 'chamber.nox_mol_mol', 'inlet.composition', | |
| 'calibration.zero_offset_no_mol_mol', 'calibration.zero_offset_nox_mol_mol'): | |
| REFERENCE_ZH[_k] = '摩尔分数(湿基:样气按到达状态,不干燥;1 ppm = 1e-6 mol/mol),与温压无关。' | |
| REFERENCE_ZH['line.condensed_water_fraction'] = '每摩尔入口湿样气在未加热路径中冷凝出的水(摩尔比)。' | |
| def validate(schema, value): | |
| if 'anyOf' in schema: | |
| for s in schema['anyOf']: | |
| try: | |
| validate(s, value) | |
| return | |
| except ActionRejected: | |
| continue | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| typ = schema.get('type') | |
| if typ == 'number': | |
| if isinstance(value, bool) or not isinstance(value, (int, float)) or not math.isfinite(value): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| if value < schema.get('minimum', -math.inf) or value > schema.get('maximum', math.inf): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| elif typ == 'integer': | |
| if isinstance(value, bool) or not isinstance(value, int): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| if value < schema.get('minimum', -math.inf) or value > schema.get('maximum', math.inf): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| elif typ == 'boolean' and type(value) is not bool: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| elif typ == 'null' and value is not None: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| elif typ == 'string' and not isinstance(value, str): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| elif typ == 'array': | |
| if not isinstance(value, list) or len(value) < schema.get('minItems', 0) or len(value) > schema.get('maxItems', 10 ** 9): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| for v in value: | |
| validate(schema.get('items', {}), v) | |
| elif typ == 'object': | |
| if not isinstance(value, dict): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| props = schema.get('properties', {}) | |
| if not set(schema.get('required', [])) <= value.keys(): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| for key, val in value.items(): | |
| if not isinstance(key, str): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| sub = props.get(key, schema.get('additionalProperties', True)) | |
| if sub is False: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| if isinstance(sub, dict): | |
| validate(sub, val) | |
| if 'enum' in schema and value not in schema['enum']: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| if 'const' in schema and value != schema['const']: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| def _clean_composition(comp): | |
| known = set(N_ATOMS) | set(quench_coefficients()) | {'oxygen', 'nitrogen', 'water', 'ozone', 'carbon_dioxide', 'carbon_monoxide', 'methane', | |
| 'sulfur_dioxide', 'hydrogen_sulfide', 'argon', 'hydrogen'} | |
| if not set(comp) <= known: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| total = sum(comp.values()) | |
| if total > 1.0 + 1e-9: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| out = {k: float(v) for k, v in comp.items() if v > 0} | |
| out['nitrogen'] = out.get('nitrogen', 0.0) + max(0.0, 1.0 - total) # balance N2 (documented) | |
| return out | |
| class EcoNcld822Runtime: | |
| """Process model + host clock state for one analyzer instance (the host is the clock authority: step(dt_s) only).""" | |
| def __init__(self, instance_id='ncld822_01'): | |
| self.instance_id = instance_id | |
| self.time_s, self.revision = 0.0, 0 | |
| self.powered = False | |
| self.t_conv = self.t_pmt = T_REF | |
| self.o3_frac = 0.0 | |
| self.full_scales = list(DEFAULT_FS) | |
| self.range_index = 2 | |
| self.filter_tau = 1.0 | |
| self.measuring = False | |
| self.cal_phase, self.cal_op, self._cal_t, self._cal_sum = 'idle', None, 0.0, [0.0, 0.0] | |
| self.cal_last = None | |
| self.span_no = self.span_nox = 2954e-6 # SRM 2631a lot value (valid certificate) | |
| self.cal_window = 30.0 | |
| self.zero_no = self.zero_nox = 0.0 | |
| self.gain_no = self.gain_nox = 1.0 | |
| self.plant = dict(PLANT_EXAMPLE, line_length_m=0.0) | |
| self.source = 'ambient_air' | |
| self.comp = self._ambient_air() | |
| self.p_in, self.T_in = self.plant['ambient_pressure_pa'], self.plant['ambient_temperature_k'] | |
| self.p_ind = self.p_in | |
| self.wall = dict(no_ch=0.0, no=0.0, no2=0.0) # channel parts after the line wall memory | |
| self.queue = deque() # (arrival time, chamber NO, chamber NOx, quench, condensed) | |
| self.ch = [0.0, 0.0] # mixed chamber equivalents (NO, NOx), quench applied | |
| self.ch_true = [0.0, 0.0] | |
| self.cond = 0.0 | |
| self.y = [0.0, 0.0] # filtered indicated (raw units, before calibration) | |
| self.noise = [0.0, 0.0] | |
| self.series = [] | |
| self._next_sample_t = 0.0 | |
| self._rng = random.Random(f'{instance_id}:0') | |
| self._cache = {} | |
| self._operation_count = 0 | |
| self._last_event = 0.0 | |
| self.label = None | |
| self._principals = {'__public_agent__': 'agent'} | |
| self._requests, self._owners = {}, {} | |
| self._arrived = self._transport(self.comp) | |
| self.queue.append((0.0, self._arrived)) # the line is already filled with room air | |
| # -------------------------------------------------------------------------------------------------------------- helpers | |
| def _ambient_air(self): | |
| T, p, rh = self.plant['ambient_temperature_k'], self.plant['ambient_pressure_pa'], self.plant['ambient_relative_humidity'] | |
| xw = min(0.2, rh * p_sat_water(T) / p) | |
| base = {'nitrogen': 0.7808, 'oxygen': 0.2095, 'argon': 0.0093, 'carbon_dioxide': 0.0004} | |
| return _clean_composition({k: v * (1 - xw) for k, v in base.items()} | {'water': xw}) | |
| def _flow(self): | |
| return Q_NOM * self.p_in / P_REF | |
| def _path_temperature(self): | |
| return T_HOT_S if self.plant['hot_tubing_installed'] else min(self.plant['line_temperature_k'] if self.plant['line_length_m'] > 0 else 1e9, | |
| self.plant['ambient_temperature_k']) | |
| def _line_volume(self): | |
| return math.pi * self.plant['line_inner_diameter_m'] ** 2 / 4.0 * self.plant['line_length_m'] | |
| def _residence(self): | |
| q = max(self._flow(), 1e-12) | |
| return LAG_INT_S * Q_NOM / q + self._line_volume() * (self.p_in / P_REF) / q | |
| def _transport(self, comp): | |
| """Inlet composition -> (NO channel eq., NOx channel eq., quench factor, condensed water, used records); cached.""" | |
| key = json.dumps([sorted(comp.items()), self.p_in, self.T_in, sorted(self.plant.items())], sort_keys=True, default=str) | |
| if key in self._cache: | |
| return self._cache[key] | |
| used = set() | |
| x = dict(comp) | |
| cond = 0.0 | |
| T_path = self._path_temperature() | |
| xw = x.get('water', 0.0) | |
| ps = p_sat_water(T_path) | |
| if xw * self.p_in > ps: # condensation in an unheated path: Henry partition | |
| x_sat = ps / self.p_in | |
| cond = (xw - x_sat) / (1.0 - x_sat) # mol water condensed per mol wet inlet gas | |
| v_l_per_v_g = cond * (0.018015 / 997.0) / (R_GAS * T_path / self.p_in) | |
| for sid in list(x): | |
| h = henry_constant(sid, T_path) | |
| if h is None or sid in ('water',): | |
| continue | |
| f = 1.0 / (1.0 + h * R_GAS * T_path * v_l_per_v_g) | |
| x[sid] *= f | |
| used.add('henry_' + sid) | |
| x['water'] = x_sat * (1.0 - cond) | |
| tot = sum(x.values()) | |
| x = {k: v / tot for k, v in x.items()} | |
| chem_in = {s: x.get(s, 0.0) for s in CHEM} | |
| clamped = [] | |
| out = integrate_dark_chemistry(chem_in, T_path if T_path < 1e8 else self.T_in, self.p_in, self._residence(), clamped) | |
| if any(chem_in[s] > 0 for s in ('nitric_oxide', 'nitrogen_dioxide', 'ozone', 'nitrate_radical', 'dinitrogen_pentoxide')): | |
| used.update(f'kinetics_iupac_{i}' for i in range(53, 65)) | |
| for s in CHEM: | |
| x[s] = out[s] | |
| no_ch = x.get('nitric_oxide', 0.0) | |
| eff = self.plant['converter_no2_efficiency'] | |
| part = self.plant['partial_conversion_fraction'] | |
| nox_no2, nox_other, partial_used = 0.0, 0.0, False | |
| for sid, v in x.items(): | |
| if sid == 'nitric_oxide' or v <= 0 or sid not in NO_PER_MOLECULE: | |
| continue | |
| cls = converter_class(sid) | |
| if cls is None: | |
| continue | |
| used.add('converter_' + CLASS_ALIAS.get(sid, sid)) | |
| if cls == 'efficient': | |
| frac = eff if sid in ('nitrogen_dioxide', 'dinitrogen_trioxide', 'dinitrogen_tetroxide') else 1.0 | |
| elif cls == 'not_efficient': | |
| frac = 0.0 | |
| else: | |
| frac = 0.5 if part is None else part | |
| partial_used = True | |
| if sid in ('nitrogen_dioxide', 'dinitrogen_trioxide', 'dinitrogen_tetroxide'): | |
| nox_no2 += frac * NO_PER_MOLECULE[sid] * v | |
| else: | |
| nox_other += frac * NO_PER_MOLECULE[sid] * v | |
| bias = 1.0 + o2_converter_bias(x.get('oxygen', 0.0)) | |
| parts = dict(no=no_ch * bias, no2=nox_no2 * bias, other=nox_other * bias) # NOx channel = NO-, NO2- and other-NOy-derived NO | |
| nox_ch = sum(parts.values()) | |
| used.add('converter_o2_effect_stainless_800c') | |
| a = quench_coefficients(self.plant['quench_relative_scale']) | |
| qf = 1.0 / (1.0 + sum(a.get(s, 0.0) * v for s, v in x.items())) | |
| used.update({'quench_if_carbon_dioxide', 'quench_if_water', 'quench_if_oxygen', 'quench_rel_nitrogen'}) | |
| if self.plant['quench_relative_scale'] > 0: | |
| used.update('quench_rel_' + s for s in ('argon', 'hydrogen', 'nitrous_oxide', 'nitrogen_dioxide') if x.get(s, 0) > 0) | |
| res = dict(no=no_ch, nox=nox_ch, nox_parts=parts, quench=qf, cond=cond, used=sorted(used), partial=partial_used, after=x, | |
| kinetics_clamped=clamped) | |
| if len(self._cache) > 256: | |
| self._cache.clear() | |
| self._cache[key] = res | |
| return res | |
| def _ready(self): | |
| return self.powered and abs(self.t_conv - T_CONV_SET) < 10.0 and abs(self.t_pmt - T_PMT_SET) < 1.0 | |
| def _status(self): | |
| return 'off' if not self.powered else ('ready' if self._ready() else 'warming') | |
| def _indicated(self): | |
| """(NO, NOx) indicated mole fractions after calibration (None when off).""" | |
| if not self.powered: | |
| return None | |
| return [self.gain_no * (self.y[0] + self.noise[0] - self.zero_no), self.gain_nox * (self.y[1] + self.noise[1] - self.zero_nox)] | |
| def _display(self): | |
| ind = self._indicated() | |
| if ind is None: | |
| return None, None, None | |
| no, nox = (round(v / DIGITS) * DIGITS for v in ind) | |
| return no, round((nox - no) / DIGITS) * DIGITS, nox | |
| def _flags(self): | |
| f = [] | |
| if self.powered and not self._ready(): | |
| f.append('warming_up') | |
| no, no2, nox = self._display() | |
| fs = self.full_scales[self.range_index] | |
| if nox is not None and max(abs(no), abs(nox)) > fs: | |
| f.append('over_range') | |
| if not (P_MIN <= self.p_ind <= P_MAX): # the analyzer only knows its indicated inlet pressure | |
| f.append('inlet_pressure_out_of_range') | |
| if not (278.15 <= self.plant['ambient_temperature_k'] <= 313.15): | |
| f.append('ambient_out_of_range') | |
| if self.powered and self.o3_frac < 0.5: | |
| f.append('no_ozone') | |
| if self.cal_phase != 'idle': | |
| f.append('calibration_running') | |
| return f | |
| def _values(self): | |
| no, no2, nox = self._display() | |
| return {'power.enabled': self.powered, 'status.phase': self._status(), 'status.flags': self._flags(), | |
| 'converter.temperature_k': self.t_conv, 'pmt.temperature_k': self.t_pmt, 'ozone.generator_fraction': self.o3_frac, | |
| 'range.full_scales_mol_mol': list(self.full_scales), 'range.index': self.range_index, | |
| 'range.active_full_scale_mol_mol': self.full_scales[self.range_index], 'filter.time_constant_s': self.filter_tau, | |
| 'measurement.phase': 'measuring' if self.measuring else 'idle', 'measurement.points': len(self.series), | |
| 'calibration.phase': self.cal_phase, 'calibration.span_no_mol_mol': self.span_no, 'calibration.span_nox_mol_mol': self.span_nox, | |
| 'calibration.window_s': self.cal_window, 'calibration.zero_offset_no_mol_mol': self.zero_no, | |
| 'calibration.zero_offset_nox_mol_mol': self.zero_nox, 'calibration.gain_no': self.gain_no, 'calibration.gain_nox': self.gain_nox, | |
| 'inlet.source': self.source, 'inlet.indicated_pressure_pa': float(round(self.p_ind)), | |
| 'flow.indicated_sample_m3_s': self._flow() if self.powered else 0.0, 'inlet.composition': dict(self.comp), 'inlet.pressure_pa': self.p_in, | |
| 'inlet.temperature_k': self.T_in, 'chamber.no_mol_mol': self.ch_true[0], 'chamber.nox_mol_mol': self.ch_true[1], | |
| 'line.condensed_water_fraction': self._arrived['cond'], 'inlet.label': self.label, | |
| 'converter.partial_assumption_applied': bool(self._arrived.get('partial') and self.plant['partial_conversion_fraction'] is None), 'display.no_mol_mol': no, 'display.no2_mol_mol': no2, 'display.nox_mol_mol': nox} | |
| def read(self, visibility='agent'): | |
| return copy.deepcopy({k: v for k, v in self._values().items() if SLOTS[k][3] == 'agent' or visibility in ('evaluator', 'debug')}) | |
| # -------------------------------------------------------------------------------------------------------------- protocol | |
| def register_principal(self, principal_id, visibility='agent'): | |
| if visibility not in ('agent', 'evaluator', 'debug'): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| if principal_id in self._principals: | |
| raise ActionRejected('REQUEST_CONFLICT') | |
| self._principals[principal_id] = visibility | |
| def session(self, principal_id='__public_agent__'): | |
| if principal_id not in self._principals: | |
| raise ActionRejected('ACCESS_DENIED') | |
| return EcoNcld822Session(self, principal_id) | |
| def describe(self): | |
| return self.session().describe(instance_id=self.instance_id) | |
| def debug_snapshot(self): | |
| return copy.deepcopy(dict(state=self.read('evaluator'), queue=list(self.queue), ch=self.ch, y=self.y, noise=self.noise, series=self.series, | |
| wall=self.wall, cal=(self.cal_phase, self.cal_op, self._cal_t, self._cal_sum), | |
| records_used=self._arrived['used'], kinetics_clamped=self._arrived.get('kinetics_clamped'))) | |
| def _measure(self, max_points=60): | |
| no, no2, nox = self._display() | |
| flags = self._flags() | |
| quality = 'off' if not self.powered else ('valid' if not flags else 'invalid') | |
| lib = library() | |
| ppm = (lambda v: None if v is None else round(v / PPM, 1)) | |
| return dict(status=self._status(), measurement='measuring' if self.measuring else 'idle', quality=quality, | |
| no_mol_mol=no, no2_mol_mol=no2, nox_mol_mol=nox, no_ppm=ppm(no), no2_ppm=ppm(no2), nox_ppm=ppm(nox), | |
| range_index=self.range_index, full_scale_mol_mol=self.full_scales[self.range_index], flags=flags, | |
| series=copy.deepcopy(self.series[-max_points:] if max_points else []), template_source=TEMPLATE_REL, library_sha256=lib['sha256'], | |
| template_records=MODEL_RECORDS, evidence='simulated_from_cited_response_parameters') | |
| def invoke(self, capability, **a): | |
| if capability in UNSUPPORTED: | |
| raise ActionRejected('UNSUPPORTED_OPTION') | |
| if capability not in CAPS: | |
| raise ActionRejected('UNKNOWN_CAPABILITY') | |
| validate(CAPS[capability][0], a) | |
| if capability == 'nox.measure': | |
| return self._measure(a.get('max_points', 60)) | |
| if capability in ('range.configure', 'calibration.configure', 'plant.configure') and self.cal_phase != 'idle': | |
| raise ActionRejected('BUSY') | |
| if capability == 'power.set' or capability == 'power.toggle': | |
| on = a['enabled'] if capability == 'power.set' else not self.powered | |
| self.powered = on | |
| if not on: | |
| self.measuring = False | |
| if self.cal_phase != 'idle': | |
| self.cal_phase, self.cal_last = 'idle', dict(operation_id=self.cal_op, status='cancelled') | |
| result = {'enabled': on} | |
| elif capability == 'range.configure': | |
| fs = [float(v) for v in a['full_scales_mol_mol']] | |
| if any(b <= c for c, b in zip(fs, fs[1:])): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| self.full_scales = fs | |
| result = {'full_scales_mol_mol': list(fs)} | |
| elif capability == 'range.select': | |
| self.range_index = a['index'] | |
| result = {'index': a['index']} | |
| elif capability == 'filter.set': | |
| self.filter_tau = float(a['time_constant_s']) | |
| result = {'time_constant_s': self.filter_tau} | |
| elif capability in ('measurement.start', 'measurement.stop', 'measurement.toggle'): | |
| start = capability == 'measurement.start' or (capability == 'measurement.toggle' and not self.measuring) | |
| if start: | |
| if not self._ready(): | |
| raise ActionRejected('PRECONDITION_FAILED') | |
| if not self.measuring: | |
| self.measuring, self.series, self._next_sample_t = True, [], self.time_s | |
| self._record() | |
| else: | |
| self.measuring = False | |
| result = {'phase': 'measuring' if self.measuring else 'idle'} | |
| elif capability == 'calibration.configure': | |
| self.span_no, self.span_nox, self.cal_window = float(a['span_no_mol_mol']), float(a['span_nox_mol_mol']), float(a['window_s']) | |
| result = dict(span_no_mol_mol=self.span_no, span_nox_mol_mol=self.span_nox, window_s=self.cal_window) | |
| elif capability in ('calibration.zero', 'calibration.span'): | |
| if self.cal_phase != 'idle': | |
| raise ActionRejected('BUSY') | |
| if not self._ready(): | |
| raise ActionRejected('PRECONDITION_FAILED') | |
| self._operation_count += 1 | |
| kind = 'zeroing' if capability == 'calibration.zero' else 'spanning' | |
| self.cal_phase, self.cal_op = kind, f'{self.instance_id}:{kind}:{self._operation_count}' | |
| self._cal_t, self._cal_sum = 0.0, [0.0, 0.0] | |
| result = {'phase': kind, 'operation_id': self.cal_op} | |
| elif capability == 'calibration.cancel': | |
| if self.cal_phase != 'idle': | |
| self.cal_phase, self.cal_last = 'idle', dict(operation_id=self.cal_op, status='cancelled') | |
| result = {'phase': self.cal_phase} | |
| elif capability in ('inlet.connect_standard', 'inlet.disconnect', 'inlet.connect'): | |
| if capability == 'inlet.connect_standard': | |
| sid = a['standard_id'] | |
| if sid in STANDARD_EXTRA: | |
| comp = _clean_composition(STANDARD_EXTRA[sid]) | |
| else: | |
| sp, xt = cylinder_truth(self.instance_id, sid) | |
| comp = _clean_composition({sp: xt, 'nitrogen': 1.0 - xt}) | |
| p, T, src = P_REF, self.plant['ambient_temperature_k'], 'standard:' + sid | |
| elif capability == 'inlet.disconnect': | |
| comp, p, T, src = self._ambient_air(), self.plant['ambient_pressure_pa'], self.plant['ambient_temperature_k'], 'ambient_air' | |
| else: | |
| if not a['label'] or len(a['label']) > 64: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| comp, p, T, src = _clean_composition(a['composition']), float(a['pressure_pa']), float(a['temperature_k']), 'sample' | |
| self.label = a['label'] | |
| if capability != 'inlet.connect': | |
| self.label = None | |
| self.comp, self.p_in, self.T_in, self.source = comp, p, T, src | |
| self._arrived = self._transport(comp) | |
| self._enqueue() | |
| if capability == 'inlet.connect_standard': | |
| if a['standard_id'] in STANDARD_EXTRA: | |
| cert = dict(certificate_mole_fraction=None, expanded_uncertainty_U95=None, value_kind='built_in_zero_gas', certificate_validity=None, | |
| certificate_expired=False) | |
| else: | |
| sp, x, u, kind = certificates()[a['standard_id']] | |
| validity = certificate_validity(a['standard_id']) | |
| cert = dict(certificate_mole_fraction=x, expanded_uncertainty_U95=u, value_kind=kind, certificate_validity=validity, | |
| certificate_expired=bool(validity and 'expired' in validity)) | |
| result = dict(standard_id=a['standard_id'], source=src, **cert) | |
| else: | |
| result = {'source': src} | |
| elif capability == 'plant.configure': | |
| new = dict(self.plant) | |
| new.update(a) | |
| if new['line_length_m'] > 0 and new['line_material'] not in ('ptfe', 'stainless'): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| self.plant = new | |
| if self.source == 'ambient_air': | |
| self.comp, self.p_in, self.T_in = self._ambient_air(), new['ambient_pressure_pa'], new['ambient_temperature_k'] | |
| self._cache.clear() | |
| if 'seed' in a: # re-seed only when a seed is given | |
| self._rng = random.Random(f'{self.instance_id}:{new["seed"]}') | |
| self._arrived = self._transport(self.comp) | |
| self._enqueue() | |
| result = {'configured': True} | |
| self.revision += 1 | |
| self._last_event = self.time_s | |
| return copy.deepcopy(result) | |
| def _enqueue(self): | |
| self.queue.append((self.time_s + self._residence(), self._arrived)) | |
| def _record(self): | |
| no, no2, nox = self._display() | |
| self.series.append(dict(time_s=self.time_s, no_mol_mol=no, no2_mol_mol=no2, nox_mol_mol=nox)) | |
| if len(self.series) > SERIES_MAX: | |
| del self.series[0] | |
| # -------------------------------------------------------------------------------------------------------------- physics | |
| def step(self, dt_s): | |
| if isinstance(dt_s, bool) or not isinstance(dt_s, (int, float)) or not math.isfinite(dt_s) or dt_s <= 0: | |
| raise ValueError('dt_s must be finite and positive') | |
| left = float(dt_s) | |
| while left > 1e-12: | |
| quiet = self.time_s - self._last_event > 30.0 # all first-order updates are exact for any h; only the | |
| h = min(left, 0.5 if quiet else 0.05) # arrival of the delayed gas needs fine steps after a change | |
| self._advance(h) | |
| left -= h | |
| self.revision += 1 | |
| return self.read('evaluator') | |
| def _advance(self, h): | |
| t1 = self.time_s + h | |
| amb = self.plant['ambient_temperature_k'] | |
| tc, tp = (T_CONV_SET, T_PMT_SET) if self.powered else (amb, amb) | |
| self.t_conv = tc + (self.t_conv - tc) * math.exp(-h / TAU_CONV_S) | |
| self.t_pmt = tp + (self.t_pmt - tp) * math.exp(-h / TAU_PMT_S) | |
| o3t = 1.0 if self.powered else 0.0 | |
| self.o3_frac = o3t + (self.o3_frac - o3t) * math.exp(-h / TAU_O3_S) | |
| # inlet pressure sensor (compensation input); option r regulates the chamber feed, so no transient error | |
| self.p_ind = self.p_in + (self.p_ind - self.p_in) * math.exp(-h / TAU_P_SENSOR_S) | |
| # delayed arrival at the chambers, then the external-line wall memory (NASA TM X-3229 rise / decay times per wetted area) on the | |
| # NO-derived and NO2-derived parts of each channel (linear filters commute with the plug-flow delay; species without wall data pass) | |
| while len(self.queue) > 1 and self.queue[1][0] <= t1: | |
| self.queue.popleft() | |
| arr = self.queue[0][1] if self.queue and self.queue[0][0] <= t1 else None | |
| area = math.pi * self.plant['line_inner_diameter_m'] * self.plant['line_length_m'] | |
| raw = dict(no_ch=0.0, no=0.0, no2=0.0, other=0.0) if arr is None else dict(no_ch=arr['no'], **arr['nox_parts']) | |
| for key, sid in (('no_ch', 'nitric_oxide'), ('no', 'nitric_oxide'), ('no2', 'nitrogen_dioxide')): | |
| tr, td = adsorption_tau(sid, self.plant['line_material'], area) | |
| tau = tr if raw[key] > self.wall[key] else td | |
| self.wall[key] = raw[key] if tau <= 0 else raw[key] + (self.wall[key] - raw[key]) * math.exp(-h / tau) | |
| truth = [self.wall['no_ch'], self.wall['no'] + self.wall['no2'] + raw['other']] | |
| if arr is not None: | |
| p_comp = 1.0 if self.plant['pressure_regulation_installed'] else (self.p_in / max(self.p_ind, 1.0)) | |
| target = [v * arr['quench'] * self.o3_frac * p_comp for v in truth] | |
| else: | |
| target = [0.0, 0.0] | |
| a_mix = 1.0 - math.exp(-h / TAU_MIX_S) | |
| for c in (0, 1): | |
| self.ch[c] += (target[c] - self.ch[c]) * a_mix | |
| self.ch_true[c] += (truth[c] - self.ch_true[c]) * a_mix | |
| # output filter and correlated zero noise (Ornstein-Uhlenbeck with the filter's correlation time) | |
| tau_f = self.filter_tau | |
| a_f = 1.0 if tau_f <= 0 else 1.0 - math.exp(-h / tau_f) | |
| tau_n = max(tau_f, 0.1) | |
| sig = NOISE0 * math.sqrt(FILTER_REF_S / tau_n) if (self.plant['noise_enabled'] and self.powered) else 0.0 | |
| rho = math.exp(-h / tau_n) | |
| for c in (0, 1): | |
| self.y[c] += (self.ch[c] - self.y[c]) * a_f | |
| self.noise[c] = self.noise[c] * rho + sig * math.sqrt(max(0.0, 1.0 - rho * rho)) * self._rng.gauss(0.0, 1.0) | |
| self.time_s = t1 | |
| # calibration averaging | |
| if self.cal_phase != 'idle' and self.powered: | |
| for c in (0, 1): | |
| self._cal_sum[c] += (self.y[c] + self.noise[c]) * h | |
| self._cal_t += h | |
| if self._cal_t >= self.cal_window - 1e-9: | |
| m = [s / self._cal_t for s in self._cal_sum] | |
| if self.cal_phase == 'zeroing': | |
| self.zero_no, self.zero_nox = m | |
| self.cal_last = dict(operation_id=self.cal_op, status='succeeded', kind='zero', mean_raw=m) | |
| else: | |
| resp = [m[0] - self.zero_no, m[1] - self.zero_nox] | |
| if resp[0] < 0.2 * self.span_no or resp[1] < 0.2 * self.span_nox: | |
| self.cal_last = dict(operation_id=self.cal_op, status='failed', kind='span', reason='SPAN_RESPONSE_TOO_LOW') | |
| else: | |
| self.gain_no, self.gain_nox = self.span_no / resp[0], self.span_nox / resp[1] | |
| self.cal_last = dict(operation_id=self.cal_op, status='succeeded', kind='span', mean_raw=m) | |
| self.cal_phase = 'idle' | |
| if self.measuring: | |
| while self._next_sample_t + SERIES_DT <= self.time_s + 1e-9: | |
| self._next_sample_t += SERIES_DT | |
| self._record() | |
| class EcoNcld822Session: | |
| """Identity-filtered protocol facade (registered principals; public calls have no visibility override).""" | |
| def __init__(self, host, principal_id): | |
| self._host, self._principal = host, principal_id | |
| def _role(self): | |
| return self._host._principals[self._principal] | |
| def _target(self, i): | |
| if i != self._host.instance_id: | |
| raise ActionRejected('UNKNOWN_INSTANCE') | |
| def describe(self, *, instance_id=None, asset_type_id=None): | |
| if (instance_id is None) == (asset_type_id is None): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| if instance_id is not None: | |
| self._target(instance_id) | |
| if asset_type_id is not None and asset_type_id != ASSET_TYPE_ID: | |
| raise ActionRejected('UNKNOWN_INSTANCE') | |
| meta = export_eco_ncld822_contract_metadata() | |
| return dict(protocol_version=PROTOCOL_VERSION, asset_type_id=ASSET_TYPE_ID, asset_version='1.0.0', | |
| capabilities={k: v for k, v in meta['capabilities'].items() if v['visibility'] == 'agent' or self._role != 'agent'}, | |
| state_slots={k: v for k, v in meta['state_slots'].items() if v['visibility'] == 'agent' or self._role != 'agent'}) | |
| def read(self, *, instance_id, state_ids=None): | |
| self._target(instance_id) | |
| h = self._host | |
| vals = h.read(self._role) | |
| ids = list(vals) if state_ids is None else state_ids | |
| if not isinstance(ids, list) or any(not isinstance(k, str) for k in ids): | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| if any(k not in vals for k in ids): | |
| raise ActionRejected('ACCESS_DENIED') | |
| quality = {'derived': 'estimated', 'setpoint': 'valid', 'process_truth': 'valid'} | |
| return dict(status='succeeded', instance_id=instance_id, state_revision=h.revision, clock={'domain': 'simulation', 'time_s': h.time_s}, | |
| observations={k: dict(value=copy.deepcopy(vals[k]), unit=SLOTS[k][1], quality=quality.get(SLOTS[k][2], 'valid'), | |
| sample_time_s=h.time_s, sample_window_s=None) for k in ids}) | |
| def invoke(self, *, instance_id, capability_id, arguments, request_id): | |
| try: | |
| return self._invoke(instance_id=instance_id, capability_id=capability_id, arguments=arguments, request_id=request_id) | |
| except ActionRejected as exc: | |
| return dict(protocol_version=PROTOCOL_VERSION, instance_id=instance_id, request_id=request_id, operation_id=None, status='rejected', | |
| state_revision=self._host.revision, clock={'domain': 'simulation', 'time_s': self._host.time_s}, result={}, observations={}, | |
| error=dict(code=str(exc), message=str(exc), retryable=False, side_effects='none')) | |
| def _invoke(self, *, instance_id, capability_id, arguments, request_id): | |
| self._target(instance_id) | |
| h = self._host | |
| if not isinstance(capability_id, str) or (capability_id not in CAPS and capability_id not in UNSUPPORTED): | |
| raise ActionRejected('UNKNOWN_CAPABILITY') | |
| if capability_id in UNSUPPORTED: | |
| raise ActionRejected('UNSUPPORTED_OPTION') | |
| if self._role == 'agent' and CAPS[capability_id][5] != 'agent': | |
| raise ActionRejected('ACCESS_DENIED') | |
| if not isinstance(request_id, str) or not request_id: | |
| raise ActionRejected('INVALID_ARGUMENT') | |
| validate(CAPS[capability_id][0], arguments) | |
| key = (self._principal, instance_id, request_id) | |
| digest = json.dumps([capability_id, arguments], sort_keys=True, allow_nan=False) | |
| if key in h._requests: | |
| old, result = h._requests[key] | |
| if old != digest: | |
| raise ActionRejected('REQUEST_CONFLICT') | |
| return copy.deepcopy(result) | |
| result = h.invoke(capability_id, **arguments) | |
| op = result.get('operation_id') if isinstance(result, dict) else None | |
| if op: | |
| h._owners[op] = self._principal | |
| response = dict(protocol_version=PROTOCOL_VERSION, instance_id=instance_id, request_id=request_id, operation_id=op, | |
| status='accepted' if op else 'succeeded', state_revision=h.revision, clock={'domain': 'simulation', 'time_s': h.time_s}, | |
| result=result, observations={}, error=None) | |
| h._requests[key] = (digest, copy.deepcopy(response)) | |
| return response | |
| def write(self, *, instance_id, state_id, value, request_id): | |
| self._target(instance_id) | |
| if state_id not in SLOTS or not SLOTS[state_id][5]: | |
| raise ActionRejected('ACCESS_DENIED') | |
| cap, arg = SLOTS[state_id][5] | |
| return self.invoke(instance_id=instance_id, capability_id=cap, arguments={arg: value}, request_id=request_id) | |
| def poll(self, operation_id): | |
| h = self._host | |
| if h._owners.get(operation_id) != self._principal: | |
| raise ActionRejected('ACCESS_DENIED') | |
| if operation_id == h.cal_op and h.cal_phase != 'idle': | |
| return dict(operation_id=operation_id, status='running', result=dict(phase=h.cal_phase, elapsed_s=h._cal_t)) | |
| last = h.cal_last if (h.cal_last and h.cal_last.get('operation_id') == operation_id) else None | |
| if last is None: | |
| raise ActionRejected('ACCESS_DENIED') | |
| return dict(operation_id=operation_id, status=last['status'], result=copy.deepcopy(last)) | |
| def cancel(self, operation_id, request_id): | |
| self.poll(operation_id) | |
| r = self.invoke(instance_id=self._host.instance_id, capability_id='calibration.cancel', arguments={}, request_id=request_id) | |
| return dict(operation_id=operation_id, status=self.poll(operation_id)['status'], result=r['result']) | |
| # ------------------------------------------------------------------------------------------------------------------- contract metadata | |
| def _p(value, unit, meaning, evidence='assumption', sources=('simulation_assumption',), uncertainty=None): | |
| return dict(value=value, unit=unit, meaning_zh=meaning, evidence=evidence, source_ids=list(sources), uncertainty=uncertainty) | |
| def export_eco_ncld822_contract_metadata(): | |
| caps, slots = {}, {} | |
| for key, (schema, unit, origin, visibility, components, write, meaning) in SLOTS.items(): | |
| slots[key] = dict(component_ids=components, value_schema=copy.deepcopy(schema), unit=unit, reference_conditions=REFERENCE_ZH.get(key), | |
| meaning_zh=meaning, origin=origin, visibility=visibility, access='read_write' if write else 'read_only', | |
| read_binding=dict(backend='host', artifact_id='runtime', accessor=f'EcoNcld822Session.read(state_ids=[{key!r}])'), | |
| write_via=dict(capability_id=write[0], argument=write[1], fixed_arguments={}) if write else None, | |
| sampling=dict(mode='instant', time_domain='simulation', window_description=None), | |
| source_ids=['manufacturer_primary', 'simulation_assumption'] if origin != 'process_truth' else ['simulation_assumption']) | |
| conditions = { | |
| 'measurement.start': ('预热完成(转换器、冷却PMT到位)', ['status.phase'], 'status.phase == ready'), | |
| 'measurement.toggle': ('开始时要求预热完成;停止无前提', ['status.phase', 'measurement.phase'], 'measuring or status.phase == ready'), | |
| 'calibration.zero': ('预热完成且无校准在进行;入口应接零气(运行层不检查气体身份)', ['status.phase', 'calibration.phase'], 'ready and calibration.phase == idle'), | |
| 'calibration.span': ('预热完成且无校准在进行;窗口平均响应须≥0.2×设定跨度,否则失败且不改增益', ['status.phase', 'calibration.phase'], | |
| 'ready and calibration.phase == idle'), | |
| 'range.configure': ('四值升序且在5–5000 ppm;校准中锁定', ['calibration.phase'], 'strictly ascending and calibration.phase == idle'), | |
| 'calibration.configure': ('校准中锁定', ['calibration.phase'], 'calibration.phase == idle'), | |
| 'plant.configure': ('校准中锁定', ['calibration.phase'], 'calibration.phase == idle'), | |
| 'inlet.connect': ('组成键须为模板库/基质已知物种,总和≤1(余量补N2)', [], 'known species and sum <= 1'), | |
| } | |
| for key, (inp, out, components, meaning, kind, vis, example, effects) in CAPS.items(): | |
| async_ = key in ('calibration.zero', 'calibration.span') | |
| pre = [] | |
| if key in conditions: | |
| d, states, check = conditions[key] | |
| pre = [dict(description_zh=d, state_ids=states, enforced_by='runtime', check=check)] | |
| src = ['manufacturer_primary', 'simulation_assumption'] | |
| if key in ('nox.measure', 'inlet.connect', 'inlet.connect_standard', 'calibration.span'): | |
| src += ['template_library'] | |
| caps[key] = dict(function_zh=meaning, kind=kind, visibility=vis, component_ids=components, availability='implemented', | |
| input_schema=copy.deepcopy(inp), output_schema=copy.deepcopy(out), | |
| implementation=dict(process='implemented', scene='absent', robot='absent', hardware='absent', host='implemented'), | |
| binding=dict(backend='host', artifact_id='runtime', entrypoint='asset_runtime.eco_ncld822.EcoNcld822Session.invoke', | |
| invocation=f'session.invoke(instance_id="ncld822_01", capability_id={key!r}, arguments={example!r}, request_id="unique")'), | |
| preconditions=pre, state_effects=[dict(state_id=k, timing='on_completion' if async_ else 'immediate', description_zh=meaning) | |
| for k in effects], | |
| observes=[], model_effects=dict(mode='none', description_zh='过程能力;屏幕读数与按键联动由Matterix桥接另行验证。', binding_ids=[]), | |
| execution=dict(mode='asynchronous' if async_ else 'synchronous', time_domain='simulation', | |
| completion_criterion='calibration.phase回到idle;poll给出succeeded/failed/cancelled' if async_ else | |
| '设定写入或观测返回;不推进物理时钟', | |
| completion_state_ids=['calibration.phase'] if async_ else effects, timeout_s=None, | |
| cancellation='supported' if async_ else 'unsupported', retry='same_request_id', | |
| resources=['target_instance', 'principal_request_cache'], | |
| partial_failure_zh='拒绝不改状态;取消/断电保留原校准系数;去重只在当前进程内存有效。'), | |
| errors=[dict(code=e, meaning_zh=e, side_effects='none') for e in | |
| ['UNKNOWN_INSTANCE', 'UNKNOWN_CAPABILITY', 'ACCESS_DENIED', 'INVALID_ARGUMENT', 'PRECONDITION_FAILED', 'BUSY', | |
| 'REQUEST_CONFLICT']], | |
| example_arguments=copy.deepcopy(example), verification=[], | |
| limitations=['工程过程模型,无真实硬件连接;进程内身份过滤不能替代部署隔离。'], source_ids=src) | |
| for key, (meaning, components) in UNSUPPORTED.items(): | |
| caps[key] = dict(function_zh=meaning, kind='device_command', visibility='agent', component_ids=components, availability='unsupported', | |
| input_schema=obj({}), output_schema=obj({}), | |
| implementation=dict(process='absent', scene='absent', robot='absent', hardware='absent', host='absent'), | |
| binding=None, preconditions=[], state_effects=[], observes=[], | |
| model_effects=dict(mode='none', description_zh='未实现', binding_ids=[]), | |
| execution=dict(mode='synchronous', time_domain='simulation', completion_criterion='始终返回UNSUPPORTED_OPTION', completion_state_ids=[], | |
| timeout_s=None, cancellation='unsupported', retry='never_automatic', resources=[], partial_failure_zh='无副作用'), | |
| errors=[dict(code='UNSUPPORTED_OPTION', meaning_zh='选件或接口未实现', side_effects='none')], example_arguments={}, | |
| verification=[], limitations=[meaning], source_ids=['manufacturer_primary']) | |
| lib = 'template_library' | |
| mechanisms = { | |
| 'chemiluminescence_detection': dict( | |
| component_ids=['detector_block', 'ozone_generator', 'converter', 'display_digits'], | |
| state_ids=['chamber.no_mol_mol', 'chamber.nox_mol_mol', 'display.no_mol_mol', 'display.no2_mol_mol', 'display.nox_mol_mol'], | |
| meaning_zh='双反应室:NO通道直接、NOx通道经金属转换器;NO+O3化学发光信号按第三体猝灭校正前的比例给出,读数=增益×(滤波信号+噪声−零点),NO2=NOx−NO', | |
| implementation='lumped', | |
| equations=['S_c = x_c,eq * Q_f * O3_frac * p_in / p_ind (option r: p_in / p_ind = 1)', 'Q_f = 1 / (1 + sum_i a_i x_i)', | |
| 'a_CO2 = 0.60, a_H2O = 2.1, a_O2 = 0 (measured interference factors); a_M = phi (q_M / q_N2 - 1) for Ar, H2, N2O, NO2 (phi host assumption, default 0)', | |
| 'x_NOx,eq = x_NO + eta_NO2 * x_NO2 + sum_j f_class(j) * nu_j * x_j; f: efficient 1, not_efficient 0, partial host assumption', | |
| 'x_NOx,eq *= 1 + bias_O2(x_O2) (stainless 800 C table)', | |
| 'reading_c = gain_c * (y_c + n_c - z_c); y: first-order filter tau_f; n: OU noise sigma = 0.06 ppm * sqrt(1 s / max(tau_f, 0.1 s))'], | |
| parameters={'noise_zero_1sigma': _p(NOISE0, 'mol/mol', '零点噪声1σ(厂家0.06 ppm,依滤波设定)', 'documented', ['manufacturer_primary']), | |
| 'min_detectable': _p(0.12 * PPM, 'mol/mol', '最低可检浓度(厂家0.12 ppm=2σ,依滤波设定)', 'documented', ['manufacturer_primary']), | |
| 'noise_reference_filter_s': _p(FILTER_REF_S, 's', '厂家噪声值对应的滤波时间常数(厂家未给,假设1 s)'), | |
| 'full_scale_limits': _p([FS_MIN, FS_MAX], 'mol/mol', '四个满量程可选5–5000 ppm', 'documented', ['manufacturer_primary']), | |
| 'display_resolution': _p(DIGITS, 'mol/mol', 'GUI一位小数ppm(手册截图)', 'documented', ['manufacturer_primary']), | |
| 'quench_co2': _p(0.60, '1', 'CO2干扰因子 −0.60 %/%(NASA TM X-3229)', 'documented', [lib]), | |
| 'quench_h2o': _p(2.1, '1', 'H2O干扰因子 −2.1 %/%(NASA TM X-3229)', 'documented', [lib]), | |
| 'converter_no2_efficiency_default': _p(1.0, '1', 'NO2转换效率缺省1.0(Dunlea 2007 “efficiently reduce”类别映射;宿主可改)'), | |
| 'partial_conversion_default': _p(0.5, '1', '“部分还原”类(HONO、NH3、PAN、HO2NO2、颗粒硝酸盐)缺省0.5为显式假设,无文献数值;宿主可改'), | |
| 'quench_relative_scale_default': _p(0.0, '1', '相对猝灭效率的换算系数φ(样气占反应室第三体的份额)缺省0:同一来源实测O2干扰≈0与φ=1矛盾;宿主可设'), | |
| 'converter_temperature_set': _p(T_CONV_SET, 'K', '金属转换器设定温度(厂家未给,假设375 °C)'), | |
| 'pmt_temperature_set': _p(T_PMT_SET, 'K', '冷却PMT设定温度(厂家未给,假设−5 °C)')}, | |
| source_ids=['manufacturer_primary', lib, 'simulation_assumption'], | |
| limitations=['Ar/H2/N2O/NO2的猝灭只在宿主给定φ>0时生效(缺省只用CO2、H2O、O2实测干扰因子);转换器材料与温度未公开,借用钼转换器分类与不锈钢转换器氧效应表;无PMT增益漂移与故障概率。']), | |
| 'sample_transport': dict( | |
| component_ids=['sample_inlet', 'inlet_filter', 'flow_restrictors', 'vacuum_pump'], | |
| state_ids=['inlet.composition', 'inlet.pressure_pa', 'flow.indicated_sample_m3_s', 'line.condensed_water_fraction', 'chamber.no_mol_mol'], | |
| meaning_zh='入口→(外部样气管)→过滤与限流→两反应室:临界限流流量∝入口绝压;内部滞后2.0 s、混合τ0.35 s;样气管壁吸附记忆(NO、NO2,作为延迟之后的一阶滤波,按通道中NO与NO2来源部分分别取上升/衰减时间);未加热路径冷凝的亨利定律洗除;停留时间内的暗反应(IUPAC 53–64,N原子守恒,温度夹在各反应评估范围内);标准气瓶真值=证书值±U95/2的确定性偏离', | |
| implementation='lumped', | |
| equations=['Q = 1.0 L/min * p_in / 101325 Pa', 't_res = 2.0 s * Q_nom / Q + V_line * (p_in / p_ref) / Q', | |
| 'tau_wall = (t/A)_record * A_wetted / ln(1000) (rise for increasing, decay for decreasing), applied after the delay to the NO- and NO2-derived channel parts', | |
| 'x_cond = (x_w - p_sat(T_path)/p) / (1 - p_sat/p); f_i = 1 / (1 + Hscp_i(T) R T V_l / V_g)', | |
| 'dn/dt = S r(n), r from IUPAC rows 53-64 (Troe falloff with [M] = p / kT), backward Euler + Newton'], | |
| parameters={'sample_flow_nominal': _p(Q_NOM, 'm^3/s', '样气流量1.0 l/min(厂家;参考状态假设101325 Pa/298.15 K)', 'documented', ['manufacturer_primary']), | |
| 'inlet_pressure_range': _p([P_MIN, P_MAX], 'Pa', '进口绝压600–1200 mbar(厂家)', 'documented', ['manufacturer_primary']), | |
| 'lag_time_bound': _p(3.0, 's', '厂家滞后<3 s(上界)', 'documented', ['manufacturer_primary']), | |
| 'rise_time_bound': _p(1.0, 's', '厂家0–90%上升<1 s(上界)', 'documented', ['manufacturer_primary']), | |
| 'internal_lag': _p(LAG_INT_S, 's', '内部传输滞后(假设,满足<3 s)'), | |
| 'mixing_tau': _p(TAU_MIX_S, 's', '混合时间常数(假设;无滤波时0–90% 0.81 s)'), | |
| 'hot_tubing_temperature': _p(T_HOT_S, 'K', 'h选件加热管路温度(厂家未给,假设180 °C)')}, | |
| source_ids=['manufacturer_primary', lib, 'simulation_assumption'], | |
| limitations=['单级平衡冷凝洗除(真实管路有累积冷凝液与传质限制);管壁记忆只对NO、NO2有数据;外部样气管由宿主配置。']), | |
| 'warmup': dict( | |
| component_ids=['converter', 'detector_block', 'ozone_generator'], state_ids=['converter.temperature_k', 'pmt.temperature_k', 'ozone.generator_fraction', | |
| 'status.phase'], | |
| meaning_zh='上电后转换器、冷却PMT、臭氧发生器一阶趋近设定(时间常数为假设);转换器±10 K、PMT±1 K内为ready', | |
| implementation='lumped', equations=['T_next = T_set + (T - T_set) exp(-dt/tau)', 'O3_next = 1 + (O3 - 1) exp(-dt/20 s)'], | |
| parameters={'converter_tau': _p(TAU_CONV_S, 's', '转换器预热时间常数(假设)'), 'pmt_tau': _p(TAU_PMT_S, 's', 'PMT冷却时间常数(假设)'), | |
| 'ozone_tau': _p(TAU_O3_S, 's', '臭氧发生器响应(假设)')}, | |
| source_ids=['simulation_assumption'], limitations=['厂家未给预热时间;温度不对外显示。']), | |
| } | |
| _ann.annotate_export('eco_ncld822', caps, slots, mechanisms, inputs={'plant.configure': (PLANT, PLANT_EXAMPLE), 'inlet.connect': (INLET, INLET_EXAMPLE)}) | |
| return dict(asset_type_id=ASSET_TYPE_ID, catalog_id=CATALOG_ID, capabilities=caps, state_slots=slots, mechanisms=mechanisms) | |