musclemap-mlx / technical.md
josand's picture
Provide public installation and qualification documentation
a80858b verified
|
Raw History Blame Contribute Delete
8.04 kB

MuscleMap v1.4 implementation and qualification

Reference contract

The source reference is MuscleMap/MuscleMap at 35a2138, with the contrast-agnostic wholebody v1.4 checkpoint and JSON configuration from Zenodo record 21929873. Both source and publisher assets carry the MIT grant. Attribution and the complete license travel with the converted bundle. Conversion uses torch.load(weights_only=True) and the standard safetensors writer, followed by exact tensor comparisons.

The network is MONAI's 2D residual UNet: one input, 114 outputs, channels (64, 128, 256, 512, 1024), strides (2, 2, 2, 2), two residual units, instance normalization with epsilon 1e-5 and no affine parameters, LeakyReLU slope 0.01, and the upstream residual and transposed-convolution ordering. All 46 float32 checkpoint tensors retain their original names and values. The complete publisher label table is checked against the runtime: all 113 IDs, names, sides, regions and channel order must match. Background is channel/ID 0; anatomical IDs are not renumbered in the saved mask.

Preprocessing follows mm_segment.py: apply NIfTI scaling, orient to RAS, resample to [1, 1, -1] mm (retain source slice spacing), normalize nonzero voxels, crop positive foreground with margin 20, then pad at the end to at least 256x256x1. The slice inferer uses 256x256 windows, batch size 1, Gaussian sigma scale 0.125 and 90% overlap. Inverse pad/crop, bilinear spacing restoration and orientation run on logits before argmax and anatomical remapping. Largest 6-connected components are selected after assembling all chunks.

Chunking follows mm_util.run_inference: slice the last source storage axis, write each float32 chunk using the source NIfTI header, then preprocess each chunk independently. NiBabel rescales float32 data stored with an integer header; this can change zeros and therefore nonzero normalization. The native runtime reproduces that round trip using a private temporary directory. Skipping it initially caused 228 CT and 190 MRI final-voxel differences in these examples. The default is eight source slices; a chunk size at least the depth skips chunk serialization and processes the whole image. Comparisons require the same chunk size.

Network operations run through MLX on Metal in float32. NIfTI I/O, geometry, Gaussian accumulation, argmax and connected components run on the host. An unavailable Metal device fails explicitly. Stored output is int16, with the original affine, distinct qform/sform matrices and codes, and unit metadata. Unknown spatial units are interpreted as mm while retaining the original unit field; explicitly different units are rejected.

Real inputs and acceptance criteria

The complete compact CT and MRI test volumes are from TotalSegmentator at d89e47f. Its official test commands supply the CT input to CT inference and the MRI input to total_mr; modality comes from that source definition. These are real images supplied as compact test examples, rather than newly synthesized images. Their NIfTI affines define 3 mm voxels; both have unspecified unit fields, qform code 0 and sform code 2. No header fields were altered for the experiment.

Input Shape Orientation Foreground IDs present
example_ct_sm.nii.gz 122x101x30 RAS 10
example_mr_sm.nii.gz 117x91x20 LPS 10

Acceptance was fixed before comparison: finite float32 network and restored logits within atol=5e-4, rtol=1e-5; exact final anatomical masks after component filtering; equal output shape, affine, spatial forms/codes and units; int16 storage. The same tensor windows are used for network comparisons. Restored-logit comparisons include native preprocessing, Gaussian aggregation and inverse geometry. CPU reduction rounding is reported separately.

Results

Qualification on 2026-11-13 used Python 3.13, MLX 0.32.4 and float32 on an Apple Silicon Mac. The independent reference used Python 3.11, PyTorch 2.4.1, MONAI 1.3.2, NumPy 1.24.4, NiBabel 5.2.1 and SciPy 1.11.4, with one CPU thread. Reference execution completed both on the Mac and on a DGX Spark GB10 using its CPU with the same pinned dependencies. Both examples used eight-slice chunks and 90% overlap. The table reports the largest error across these two independent reference executions.

Measurement CT MRI
Maximum network-logit absolute error 1.47e-4 1.36e-4
Maximum restored-logit absolute error 1.15e-4 1.03e-4
Final mismatched voxels 0 / 369,660 0 / 212,940
Dice for every foreground ID present 1.0 1.0
Output grid, forms and units Preserved Preserved

CT preprocessing differed by at most 4.77e-7 from the reference after float32 normalization; MRI preprocessing matched exactly. Reference reader pixels were checked independently with pinned NiBabel. The complete public package runner also produced the exact reference masks. These measurements establish parity on two compact examples, without measuring manual-label accuracy or every one of the 113 anatomical structures.

Reproduce

Install the published runtime and download the public qualification tools at their reviewed revision. The reference uses Python 3.11; native execution uses Python 3.13 on Apple Silicon. The source distribution supplies the configuration parser for the reference environment, without installing an MLX runtime there.

uv venv --python 3.13
source .venv/bin/activate
uv pip install musclemap-mlx==0.1.1
hf download MedMLX/musclemap-mlx --revision 05fa11f1f13ded18d2efdb7a8e1dab77bd600234 --include 'qualification/*' --local-dir validation

# Obtain the same published source used by the installed runtime.
python - <<'PYCODE'
import json, urllib.request
metadata = json.load(urllib.request.urlopen("https://pypi.org/pypi/musclemap-mlx/0.1.1/json"))
url = next(item["url"] for item in metadata["urls"] if item["packagetype"] == "sdist")
urllib.request.urlretrieve(url, "musclemap_mlx-0.1.1.tar.gz")
PYCODE
tar -xf musclemap_mlx-0.1.1.tar.gz
uv venv --python 3.11 oracle-env
uv pip install --python oracle-env/bin/python -r validation/qualification/requirements.txt
mkdir -p weights results
curl -L https://zenodo.org/records/21929873/files/contrast_agnostic_wholebody_model.pth -o weights/contrast_agnostic_wholebody_model.pth
curl -L https://zenodo.org/records/21929873/files/contrast_agnostic_wholebody_model.json -o weights/contrast_agnostic_wholebody_model.json
curl -L https://zenodo.org/records/21929873/files/LICENSE -o weights/LICENSE
curl -L https://raw.githubusercontent.com/wasserth/TotalSegmentator/d89e47f73d24f5fba732a715aeb6fdc0ae34526b/tests/reference_files/example_ct_sm.nii.gz -o ct.nii.gz
curl -L https://raw.githubusercontent.com/wasserth/TotalSegmentator/d89e47f73d24f5fba732a715aeb6fdc0ae34526b/tests/reference_files/example_mr_sm.nii.gz -o mr.nii.gz

PYTHONPATH=musclemap_mlx-0.1.1/src OMP_NUM_THREADS=1 OPENBLAS_NUM_THREADS=1 \
  oracle-env/bin/python validation/qualification/make_reference_fixtures.py \
  --assets weights --image ct.nii.gz --output results/ct-reference
musclemap-mlx download --model-dir converted
python validation/qualification/qualify_real.py \
  --reference results/ct-reference --weights converted --output results/ct-parity.json

Repeat the last reference and comparison commands with mr.nii.gz and separate MRI result paths. The recorder can run on Linux; copy its result folder to the Mac for native comparison. qualify_real.py exits unsuccessfully for logit failures or any final mask mismatch and records per-chunk errors and per-label Dice. Generated images, weights, reference arrays and run output stay outside Git.